Meteorology · Agriculture · Technical guides
Integrated weather station on campus and at the experimental farm: installation, comparison with INMET and data quality control
A teaching guide to using an integrated automatic weather station (sensor suite + console) as a research data source: how to read the datasheet (+2 °C passive-shield error, photodiode radiation at ±5% of full scale, console ET that is not FAO-56), WMO siting (heights, site class, true north, barometer elevation), logging in UTC, intercomparison with Brazil's INMET network (MBE, RMSE, Willmott's d, Camargo & Sentelhas's c index, worked example converting wind to 2 m), Allen/FAO-56 integrity tests, maintenance and Brazilian studies from UNESP Jaboticabal.

Many departments at Brazilian public universities have — or want — an integrated automatic weather station on campus, at the experimental farm or next to a field trial. It fits a project budget, goes up in an afternoon and delivers air temperature, humidity, rainfall, wind, pressure, solar radiation and UV on a single console. The question a supervisor should ask before those data go into a thesis is not “does the station work?” but “how much do I trust each variable, and how do I demonstrate it?”
This guide is for faculty, graduate students and researchers who are about to install or inherit an integrated station. We have already covered how to choose a weather station and each sensor on its own. The angle here is different: installation at a real site, intercomparison with Brazil’s INMET network and quality control (QC) of the data — the steps that turn a campus station into a citable source.
What an integrated station is and where it fits
An integrated station puts the rain collector, the radiation shield with temperature and humidity sensors and, on a separate arm, the cup anemometer and wind vane into a single sensor assembly (Davis calls it the Integrated Sensor Suite, ISS). An indoor console reads the sensors, computes derived variables and displays everything. The “Plus” model adds solar radiation and UV sensors.
It sits between two worlds:
- Modular research stations (programmable logger + reference sensors chosen one by one): higher accuracy and traceability, higher cost and more integration work.
- All-in-one compact stations (ultrasonic sensor with no moving parts, solar panel and GSM): fast installation and low maintenance, with other limitations — see compact weather station with no moving parts.
An integrated station is excellent for characterising the microclimate of an experimental site, teaching agrometeorology, supporting irrigation management and building local rainfall and temperature records. To serve as research data, however, it needs three things the catalogue cannot provide by itself: a well-chosen site, a documented comparison with a reference and a QC protocol.
What the datasheet says — and how to read each number
Taking as reference the Davis Vantage Pro2 Plus cabled weather station in the Hensistemas catalogue (ref. 6162C), the manufacturer’s datasheet (DS6152C/6162C, 2021 revision) states:
| Variable | Manufacturer specification | How to read it in research |
|---|---|---|
| Air temperature (ISS) | −40 to +65 °C; ±0.3 °C typical | This is sensor error. Shield error is something else (see below). |
| Radiation-induced error (passive shield) | +2 °C at solar noon (1040 W/m², average wind 1 m/s) | Dominates Tmax uncertainty on clear, calm days. |
| Relative humidity | 1–100%; ±2%; drift < 0.25%/year | Check that night-time maximum RH approaches 100% on dew nights. |
| Rain collector | Tipping spoon, 0.2 mm per tip, 214 cm² area | ±3% of total or ±1 tip (up to 250 mm/h). |
| Rain rate | ±5% up to 250 mm/h | Computed from the time between tips. |
| Wind speed | ±0.9 m/s or ±5% (whichever is greater) | In light wind, the fixed 0.9 m/s term dominates. |
| Wind direction | ±3° | Requires the arm to be correctly aligned to north. |
| Pressure (in the console) | 540–1100 hPa; ±1.0 hPa | The barometer sits in the console, indoors. |
| Solar radiation | 0–1800 W/m²; ±5% of full scale; cosine ±3% up to 75°; drift up to 2%/year | ±5% of 1800 W/m² = ±90 W/m² — a silicon photodiode, not a thermopile pyranometer. |
| UV index | 0–16; ±5% of full scale | Useful for teaching and public health, not for reference UV radiometry. |
| Console ET | Modified Penman (CIMIS), hourly step; greater of 0.25 mm or ±5% | Not FAO-56 Penman-Monteith ET₀ — recompute it. |
Two points deserve emphasis:
- “% of full scale” is not “% of reading”. A sensor with ±5% FS over 0–1800 W/m² can be off by ±90 W/m² at 1000 W/m² and at 200 W/m² alike. On an overcast day that is a huge relative error. For energy balance or reference ET₀, compare against a thermopile pyranometer and see the guide on pyranometers at the weather station.
- Passive-shield error is larger than sensor error. The manufacturer itself declares +2 °C at solar noon with 1 m/s wind, referenced to an aspirated shield. In the Cerrado and the semi-arid Northeast, clear calm days are exactly the days of extreme Tmax. The datasheet lists versions with a fan-aspirated shield; with the passive shield, treat Tmax as the variable most in need of intercomparison. See the guide on air temperature, humidity and radiation shields.
Installation at a real site: what matters most
The World Meteorological Organization (Guide to Instruments and Methods of Observation, WMO-No. 8) recommends temperature and humidity between 1.25 m and 2 m above representative grass, and standard wind at 10 m over open terrain. It also describes a siting classification (Annex 1.B, equivalent to ISO 19289) that scores each sensor’s exposure by distance to obstacles, artificial surfaces and heat sources. Documenting that class is cheap and adds real value to a paper.
In practice, for an integrated station:
- T/RH shield at 1.5–2 m, over grass, away from walls, asphalt, roofs and air-conditioning units. A station on a building roof measures the roof’s microclimate.
- Rain collector level and clear of obstacles: the WMO rule of thumb is a distance of at least twice the obstacle height. Trees grow — re-check the surroundings every year.
- Anemometer as high and unobstructed as possible. The anemometer cable supplied is 12 m (up to 73 m recommended, with a lower maximum displayed wind speed on long cables). Record the actual sensor height: you will need it.
- Anemometer arm aligned to true north (not magnetic). Magnetic declination exceeds 20° in several parts of Brazil — a systematic error in the wind rose.
- Radiation and UV sensors level, with no shadow from the mast or trees at any time of year (the sun’s azimuth shifts between solstices).
- Console and power. In the cabled version the console powers the ISS through the cable (30 m included, up to 300 m recommended) and runs on a 5 V adapter with backup batteries for about a month. It is a solution for sites with mains power nearby — a building, shed or pump house. At remote sites with no grid, prefer a station with its own battery or solar panel (alternatives below).
- Elevation in the console. The barometer reduces pressure to sea level using the elevation you enter; the manufacturer asks for elevation within ±3 m. Use GNSS or a digital elevation model, not a guess. See atmospheric pressure at the weather station.
Logging: interval, clock and time zone
The console displays, but the logger keeps the record (WeatherLink, optional), with an archive interval configurable from 1 to 120 minutes. Three decisions save trouble:
- A 5 to 15 minute interval suits most agronomic and hydrological work and aggregates cleanly to hours and days. Long intervals (60 min) hide rain-intensity peaks and gusts.
- One declared time zone. Hourly data from INMET automatic stations are published in UTC. If your station logs local time (UTC−3 in most of Brazil), convert before comparing. A 3-hour offset destroys the hourly correlation of temperature and radiation and can be mistaken for “sensor error”.
- A checked clock. The console clock is accurate to ±8 s/month; over a year, that is minutes of drift. Synchronise it at every visit and note it in the field log.
For integration with other sensors and telemetry, see data logger: how to choose and weather station telemetry.
Intercomparison with the INMET network: step by step
INMET’s automatic station network is the natural reference in Brazil, and its hourly data are open on the institute’s portal. A lean protocol:
- Choose the reference station. The nearest one with similar climate and elevation. Record the distance, elevation difference and differences in surroundings (urban vs rural, nearby reservoir). Microclimate differences are not sensor error.
- Minimum period. At least a few weeks covering dry and rainy days; ideally one dry and one wet season.
- Align time and aggregation. Same time zone (UTC), same integration period (clock hour, 00–24 UTC day or local day — state which).
- Compare the right quantity. Hourly temperature and RH; daily rainfall (convective rain is extremely variable in space — the hourly correlation between points 20 km apart is naturally low); wind converted to the same height.
- Use standard statistics: mean bias error (MBE), RMSE, correlation coefficient r, Willmott’s (1981) index of agreement d and Camargo & Sentelhas’s (1997) index c = r · d, widely used in Brazilian agrometeorology (“excellent” for c > 0.85; “very good” between 0.76 and 0.85; “good” between 0.66 and 0.75).
Worked example: wind at different heights
Suppose the campus anemometer is at 3 m and records a daily mean of 2.6 m/s, while the INMET station at 10 m records 3.2 m/s. That looks like a −0.6 m/s bias. Before concluding anything, apply the logarithmic conversion from FAO-56 (eq. 47): u₂ = u_z · 4.87 / ln(67.8 z − 5.42).
- Campus, z = 3 m: factor 0.921 → u₂ = 2.6 × 0.921 ≈ 2.39 m/s
- INMET, z = 10 m: factor 0.748 → u₂ = 3.2 × 0.748 ≈ 2.39 m/s
The two stations agree: the “difference” was only height. This kind of mistake — comparing non-equivalent quantities — is more common in theses than an uncalibrated sensor.
What to do with the result
- Small, constant bias (e.g. Tmin 0.2 °C lower): document it; do not “correct” it without a physical reason.
- Radiation-dependent bias (higher Tmax on clear, calm days that vanishes on cloudy or windy days): the signature of the passive shield. Consider an aspirated shield or a T/RH reference in a multi-plate radiation shield.
- Abrupt step change in bias after a fault or maintenance: the classic problem already described by INMET researchers comparing automatic and conventional stations (Mello et al., 2006). Split the series into segments.
- Large rainfall difference in a single event but good monthly totals: spatial variability, not a defect.
Continuous quality control (without an external reference)
Even with no neighbouring station, physics allows strong checks. Allen (1996) proposed, and FAO-56 consolidated, integrity tests for data used in ET₀:
- Solar radiation vs clear-sky radiation. On clear days, Rs should approach Rso = (0.75 + 2×10⁻⁵ z) Ra (FAO-56, eq. 37). If the clearest days of the month sit systematically 10% below Rso, suspect a dirty or unlevel sensor or drift (the manufacturer declares up to 2%/year).
- Night-time maximum humidity. In humid periods, maximum RH should approach 100% on dew mornings. If it “sticks” at 92–94%, the hygrometer is degraded. This is the kind of test that led Turco & Barbosa (2008) to flag humidity problems at both stations they evaluated in Jaboticabal.
- Dew point vs Tmin. In humid climates Tmin and dew-point temperature stay close; a growing gap indicates an RH problem (or a real regime change — check).
- Rainfall. Days with tips and clear-sky radiation all day, or isolated tips with no clouds: sprinkler irrigation hitting the collector, insects or vibration.
- Wind. Zero speed for hours in a normally windy period: a seized bearing. Constant direction: a stuck vane or faulty potentiometer.
- Physical limits and jumps. Temperature outside the climatological range, jumps of several degrees between consecutive readings, identical values repeated for long periods.
Record every flag and correction in a metadata file. At the defence, “the series passed these tests and these dates were removed” is worth more than any catalogue spec.
What Brazilian research has already shown
- Turco & Barbosa (2008), in Irriga (v. 13, n. 3, p. 339–354), installed a Davis station and a Campbell Scientific station side by side over irrigated grass at FCAV/UNESP in Jaboticabal (SP), logging every 15 minutes. Daily Penman-Monteith ET₀ estimates showed acceptable differences between the two; the integrity test flagged poor-quality relative humidity data at both stations and poor rainfall data at the Campbell.
- Turco & Carleto (2017), in Revista Brasileira de Agricultura Irrigada (v. 11, n. 8, p. 2084–2097), used three automatic stations (Campbell, Davis and one from SINDA/INPE) as input to the SWAT hydrological model for the Córrego Rico basin (SP). The Campbell and Davis stations were considered usable for water-resource planning; the station without calibration, maintenance and inspection compromised the simulated hydrological parameters. The core message is methodological: without an inspection and calibration plan, no brand will save the record.
- Perazzi et al. (2021), in Revista Brasileira de Meteorologia (v. 36), compared conventional and automatic stations of the Brazilian network (2000–2019) and found good agreement in many cases but marked inconsistencies in others — a reminder that even the reference needs checking.
Maintenance: the minimum calendar
- Every visit (fortnightly or monthly): clean the rain funnel and screen, remove nests and webs from the shield, wipe the radiation and UV diffusers with a soft cloth, check levels, check the clock and download the data.
- Every 6 months: tip test with a known volume in the rain collector (pour slowly and compare with the expected value for 214 cm²), check vane alignment, inspect cables and RJ-11 connectors (moisture and rodents).
- Annually: RH and temperature intercomparison with a freshly calibrated reference sensor placed alongside for a few days; radiation drift check against a pyranometer; review of the surroundings (trees, new buildings).
- Always: a field log with date/time of every intervention — it is what explains the steps that will show up in the series.
Featured product: Davis Vantage Pro2 Plus cabled integrated station
In the Hensistemas catalogue, the Davis Vantage Pro2 Plus cabled weather station (ref. 6162C) brings together:
- wind speed and direction, outside and inside temperature and humidity, rainfall and rain rate, atmospheric pressure;
- solar radiation and UV sensors as standard (what makes it the “Plus” model), enabling ET and thermal-index calculations;
- a console with backlit LCD, on-screen graphs and more than 70 simultaneous alarms;
- a cabled link between ISS and console, with the console powering the sensor suite;
- an optional WeatherLink logger with a 1–120 minute archive interval and spreadsheet export.
For research, the strength is the cost-benefit of having all basic variables in one easy-to-install instrument that is widely used in Brazilian studies. The points to plan for are the passive shield (Tmax on clear, calm days), photodiode radiation (±5% of full scale), console ET (not FAO-56) and the need for mains power nearby in the cabled version.
Catalogue alternatives and complements
- WatchDog 2900ET weather station (ref. 3350WD2): measures and logs ET, radiation, wind, temperature, humidity and rainfall, with non-volatile memory and long battery life — a good choice where there is no mains power.
- All-in-one compact station (ref. HDMCS-200): ultrasonic wind with no moving parts, integrated solar panel, battery and GSM, for remote sites.
- Rain gauge with 400 cm² collecting surface (ref. PL400): class A tipping bucket under UNI 11452, to serve as rainfall reference beside the station; guide: tipping-bucket rain gauge.
- Relative humidity and temperature sensor with SDI-12 output (ref. SRH1A): T/RH reference for the annual intercomparison.
- Multi-plate radiation shield (ref. EIC2895-01): shield for the reference sensor.
- Kipp & Zonen pyranometer: ISO-classified thermopile, to check the photodiode and for energy-balance work.
- Data logger for weather stations (ref. HD33MT.4): when the project grows into a modular station with reference sensors and 4G telemetry.
Checklist before using the data in a thesis
- Are the site class (WMO-No. 8, Annex 1.B) and the actual height of each sensor documented?
- Is the time zone of the files stated and the same as the reference (UTC at INMET)?
- Was wind converted to the same height (FAO-56, eq. 47) before comparison?
- Is there an intercomparison period with an INMET station or side-by-side reference sensors, with MBE, RMSE, r, d and c?
- Was Tmax analysed against radiation and wind (passive-shield signature)?
- Did radiation pass the Rs vs Rso test and RH the night-time maximum test?
- Was ET₀ recomputed with FAO-56 Penman-Monteith from the measured variables instead of using console ET? See ET₀ with FAO-56.
- Is there a field log of every maintenance task and sensor swap?
References
- Davis Instruments. Cabled Vantage Pro2™ & Vantage Pro2 Plus™ Stations — Specification Sheet (DS6152C, 6162C, rev. AA, 2021).
- Hensistemas product page: Davis Vantage Pro2 Plus cabled weather station (ref. 6162C).
- World Meteorological Organization. Guide to Instruments and Methods of Observation (WMO-No. 8), Volume I, including Annex 1.B (siting classification).
- Allen, R. G.; Pereira, L. S.; Raes, D.; Smith, M. Crop evapotranspiration — Guidelines for computing crop water requirements. FAO Irrigation and Drainage Paper 56. Rome: FAO, 1998.
- Allen, R. G. Assessing integrity of weather data for reference evapotranspiration estimation. Journal of Irrigation and Drainage Engineering, v. 122, n. 2, p. 97–106, 1996.
- Turco, J. E. P.; Barbosa, J. C. Avaliação de duas estações meteorológicas automatizadas, para estimativa diária da evapotranspiração de referência obtida pelo método de Penman-Monteith. Irriga, v. 13, n. 3, p. 339–354, 2008.
- Turco, J. E. P.; Carleto, N. Integridade de dados meteorológicos para uso em modelo hidrológico. Revista Brasileira de Agricultura Irrigada, v. 11, n. 8, p. 2084–2097, 2017. DOI: 10.7127/rbai.v11n800668.
- Perazzi, P. R.; Passamani, M.; Thielen, D.; Padovani, C. O tradicional ou o moderno? Uma visão da informação da rede de estações meteorológicas brasileiras. Revista Brasileira de Meteorologia, v. 36, 2021. DOI: 10.1590/0102-77863630115.
- Mello, L. T. A.; Fortes, L. T. G.; Lucio, P. S.; Manhães, F. P.; Lucas, E. W. M. Uma análise comparativa de dados meteorológicos observados em estações automáticas e convencionais do INMET. In: XIV Congresso Brasileiro de Meteorologia, Florianópolis, 2006.
- Willmott, C. J. On the validation of models. Physical Geography, v. 2, n. 2, p. 184–194, 1981.
- Camargo, A. P.; Sentelhas, P. C. Avaliação do desempenho de diferentes métodos de estimativa da evapotranspiração potencial no Estado de São Paulo, Brasil. Revista Brasileira de Agrometeorologia, v. 5, n. 1, p. 89–97, 1997.
Instruments for this application

Davis Instruments6162C
Davis Vantage Pro2™ Plus Cabled Weather Station 6162C

Spectrum Technologies3350WD2
WatchDog 2900ET Weather Station

DeltaOhm - GHM GroupHDMCS-200
HDMCS-200 All-in-One Meteo Compact Station

NESASRLPL400
PL400 Rain Gauge with 400cm2 Collecting Surface

EIC2895-01
Multi-Plate Radiation Shield

Kipp & Zonenkipp&zonen
Kipp & Zonen Pyranometer for Solar Irradiance Measurement

DeltaOhm - GHM GroupHD33MT.4