Meteorology · Agriculture · Technical guides
Spectral light quality in plant research: how to measure the spectrum with a spectroradiometer under LEDs, coloured shade nets and greenhouses and combine it with weather-station PAR
A didactic guide to measuring spectral light quality in plant research: quantity vs. quality, how a spectroradiometer works, PPFD, ePAR (400–750 nm), band fractions, R:FR ratio and PSS calculated from the spectrum, why lux cannot equalise treatments, Brazilian examples with coloured nets and in vitro LEDs, measurement campaigns, transmittance of nets and vessels, instrument limits and combination with continuous weather-station PAR.

Anyone who works in plant physiology, horticulture, micropropagation or protected cultivation has heard the examiner's question: "but what light did the plants actually receive?". The answer usually has two parts — how much light (the photosynthetic photon flux density, PPFD, in µmol m⁻² s⁻¹) and what kind of light (the spectral quality: how much blue, green, red and far-red). The quantum sensor of the weather station or the growth chamber answers the first part well. Not the second: to know the composition of the light you need to measure the spectrum, wavelength by wavelength, with a spectroradiometer.
This guide is for lecturers, graduate students and technicians who set up experiments with LEDs, coloured (photoselective) shade nets, plastic films, in vitro culture rooms and greenhouses. It explains what a spectroradiometer measures, which indices to calculate from the spectrum, why lux cannot be used to equalise plant treatments, how to plan a measurement campaign, how to measure the transmittance of a net or a culture-vessel lid, what the instrument's limits are, and how to combine all of this with the continuous PAR from the weather station. Measuring the amount of light with a quantum sensor is covered in photosynthetically active radiation (PAR) in research.
Quantity and quality: two different questions
Plants "read" light in two ways:
- As energy for photosynthesis. What matters here is the number of photons between 400 and 700 nm — the classic definition of PAR, based on McCree's (1972) action spectra. This is what a quantum sensor measures: a single number, the PPFD.
- As a signal. Photoreceptors such as phytochromes (sensitive to red and far-red) and cryptochromes and phototropins (sensitive to blue and UV-A) tell the plant whether it is in full sun or shaded by neighbours. The response shows up in stem elongation, leaf expansion, stomatal density, pigments and secondary metabolites.
A quantum sensor is blind to the second question. Two sources with the same PPFD — a white LED and a "pink" LED (blue + red), for example — can produce plants with very different morphology. That is why, when the experimental treatment is the light, the spectrum is part of the method.
What a spectroradiometer is and how it works
A spectroradiometer splits the light reaching an input diffuser and measures the spectral irradiance, E(λ), in W m⁻² nm⁻¹, across a range of wavelengths. Typical components are:
- A cosine-corrected diffuser, so the reading represents the light reaching a flat surface (such as a leaf) from every direction of the hemisphere;
- An entrance slit and a diffraction grating, which separate the light by wavelength;
- A linear detector (CCD) with hundreds or thousands of elements, each "seeing" a narrow slice of the spectrum;
- A calibration file, obtained with a standard lamp of known spectral irradiance, which converts detector counts into W m⁻² nm⁻¹.
Three datasheet numbers say a lot about the instrument:
- Spectral range: where it "sees". For plants, ideally from 400 to at least 750 nm, to include far-red.
- Bandpass (bandwidth): the width, in nm, that each reading effectively integrates. The smaller it is, the better the instrument resolves the narrow peaks of LEDs.
- Wavelength accuracy: how far the λ axis may be shifted. An error of a few nm on the peak of a red LED changes the calculated red:far-red ratio.
Another important operating parameter is the integration time: how long the detector accumulates light. Dim light needs a long time; bright light, a short one. If the time is too long, the detector saturates and the peak of the spectrum is "clipped"; if it is too short, the spectrum becomes noisy.
The indices you calculate from the spectrum
The advantage of having E(λ) in 1 nm steps is that you can calculate any quantity afterwards — including ones that may be proposed in the future. The most common in plant research are:
PPFD (400–700 nm). Convert spectral irradiance into photon flux and sum over the band:
N(λ) [µmol m⁻² s⁻¹ nm⁻¹] = E(λ) [W m⁻² nm⁻¹] × λ [nm] ÷ 119.6
and PPFD = the sum of N(λ) from 400 to 700 nm. The 119.6 factor comes from the product of Planck's constant, the speed of light and Avogadro's number. Note that one watt of red light carries more photons than one watt of blue light: at 450 nm, 1 J corresponds to about 3.8 µmol of photons; at 660 nm, to about 5.5 µmol.
ePAR (400–750 nm). Zhen and Bugbee (2020) showed, in 14 crop species, that far-red photons (701–750 nm) added to a 400–700 nm background increased canopy photosynthesis as much as the same number of "traditional" photons — and proposed extending PAR to 750 nm. If your instrument covers this range, report PPFD and ePAR side by side, especially in experiments with far-red LEDs.
Band fractions. A common split is blue (400–500 nm), green (500–600 nm), red (600–700 nm) and far-red (700–750 nm), expressed in µmol m⁻² s⁻¹ or as a percentage of the total. This lets you write "31% blue photons" in the methods instead of "blue LED".
Red:far-red ratio (R:FR). The ratio of photon flux in a red band to that in a far-red band, close to the phytochrome absorption peaks (around 665 and 730 nm). There is no single band convention: some authors use 655–665 / 725–735 nm, others 645–685 / 710–750 nm. A technical note from Apogee Instruments, using 645–685 and 710–750 nm bands, recorded an R:FR of 1.11 in full sunlight and 0.24 in the shade beneath a canopy — the order of magnitude that separates "sun" from "shade by neighbours". State the bands in your methods, because the number changes with them.
Phytochrome photostationary state (PSS or PPE). Weights the spectrum by the absorption cross-sections of the phytochrome forms, following the methodology of Sager et al. (1988). It can only be calculated from spectral data — one of the strongest arguments for a spectroradiometer instead of two band sensors.
Photometric and colour quantities. Illuminance (lux), correlated colour temperature (CCT), chromaticity coordinates and colour rendering index (CRI) are useful in studies of indoor lighting, ergonomics and visual comfort — but they describe light as the human eye perceives it, not as the plant uses it.
Why lux cannot equalise plant treatments
Lux weights each wavelength by the sensitivity curve of the human eye, V(λ), which peaks at 555 nm and falls steeply in the blue and the red — precisely the regions where many horticultural LEDs concentrate their output. Here is an illustrative calculation with idealised narrow-band sources:
- At 450 nm, V(λ) ≈ 0.038; 1 µmol m⁻² s⁻¹ of photons corresponds to about 7 lux.
- At 660 nm, V(λ) ≈ 0.061; 1 µmol m⁻² s⁻¹ corresponds to about 7.5 lux.
- For sunlight, the classic review by Thimijan and Heins (1983) gives about 54 lux per µmol m⁻² s⁻¹.
Result: a shelf lit by blue + red LEDs and another lit by a daylight-like spectrum, both measured at "3,000 lux", may be receiving something like 415 and 56 µmol m⁻² s⁻¹, respectively — more than a seven-fold difference in photosynthetic light. The exact numbers depend on the real width of each LED's peak, but the lesson is general: a lux meter does not equalise plant treatments, and the lux → µmol conversion factor changes with the spectrum. With a measured spectrum, the conversion stops being a guess.
Something similar, on a smaller scale, happens with quantum sensors: the spectral response of a real sensor is not perfectly flat between 400 and 700 nm, and the error tends to be larger under narrow-band LEDs than under sunlight. A spectroradiometer campaign also lets you check whether the quantum sensor installed in the chamber reads correctly under that particular source.
Where spectral quality shows up in Brazilian research
A few examples show how the topic appears in Brazilian graduate programmes:
- Coloured nets. At the Federal University of Lavras (UFLA), Silva et al. (2015) grew tamarind seedlings under 50% shade nets in white, blue, red and black and found leaf anatomical plasticity and higher stomatal densities under the blue and red nets. In that work, the spectral change of each net is described according to the manufacturer. Pinto et al. (2014), in an experiment also conducted at UFLA, studied yarrow under coloured nets and measured PPFD in each environment with a portable quantum meter (roughly 850 to 950 µmol m⁻² s⁻¹ under the nets and 1,970 µmol m⁻² s⁻¹ in full sun). Measuring the spectrum under each net, on site and on the same date, complements this kind of design with the very data that define the treatment.
- LEDs in micropropagation. Silva et al. (2016), at the Federal Rural University of Pernambuco (UFRPE), compared blue:red combinations of 70:30, 50:50 and 30:70, white LEDs and fluorescent lamps for in vitro sugarcane culture; Santos Junior et al. (2022) evaluated white, red, blue and "pink" (50% blue and 50% red) LEDs for in vitro multiplication and rooting of Spanish cedar (Cedrela odorata). In designs like these, the measured spectrum makes explicit whether "50:50" refers to photons, energy or number of LEDs, and documents the peak and width of each colour.
- Protected cultivation and films. Covering films age, collect dust and change their transmittance over the years. Measuring the spectrum inside and outside the greenhouse — instead of assuming the film's catalogue data — gives the study the light that actually reached the canopy. The thermal microclimate of the greenhouse is covered in greenhouse microclimate in research.
Planning the measurement campaign
A portable spectroradiometer is a campaign instrument, not a continuous open-sky monitor. A workflow that works:
- Define the reference plane. Measure at the height of the top of the canopy (or the top of the vessels in the in vitro room), with the diffuser level and facing up. Record the distance to the source.
- Use a grid of points. Shelf lighting is rarely uniform: edges get less light than the centre. A grid of, for example, 3 × 3 points per shelf lets you report the mean and coefficient of variation of PPFD and of the spectral fractions.
- Let the source stabilise. Lamps and LEDs change slightly in intensity and spectrum in the first minutes after switching on. Note how long you waited.
- Control the integration time. Use automatic mode or adjust manually until the instrument indicates correct exposure — neither saturated nor underexposed.
- Watch out for shadows and reflections. The operator is a source of error: light-coloured clothes reflect light onto the diffuser and the body casts shade. Keep your distance and wear dark clothing.
- Repeat over time. LEDs lose output with operating hours; nets and films age. Measuring at the beginning, middle and end of the experiment documents the drift.
- Keep the raw spectrum. Export the file with E(λ) in 1 nm steps and calculate the indices later with a version-controlled script. That is what lets you recalculate ePAR, PSS or a new band ratio when the reviewer asks.
Transmittance: measuring nets, films and vessel lids
Many spectroradiometers have a spectral transmittance mode, T(λ): first a reference spectrum is measured without the material, then the spectrum with the material between the source and the diffuser. The ratio, wavelength by wavelength, is the transmittance. This is useful to:
- characterise new and aged coloured nets and covering films;
- measure how much of the culture-room light passes through the lid or wall of the culture vessel — the explant does not receive the light measured on top of the shelf, it receives the light filtered by the container;
- compare batches of material before setting up an experiment.
Practical precautions: the source must be stable between the reference and the sample reading (under a sky with passing clouds, alternate several reference and sample readings); the geometry must be the same in both readings; and materials that scatter a lot of light, such as nets, give results that depend on the distance between material and diffuser — record that distance.
The featured instrument: portable spectroradiometer for visible and ultraviolet light
For those who need this characterisation, the catalogue includes the HD30.1 spectroradiometer data logger from Delta OHM. It consists of a touch-screen indicator/logger and interchangeable probes, each with its calibration file stored inside. According to the product sheet and the manufacturer's datasheet:
- Visible probe (HD30.S1): 380 to 780 nm; calculates illuminance, CCT, chromaticity coordinates (CIE 1931 and CIE 1976), CRI and PAR; 4.5 nm bandpass; calibrated with a standard halogen lamp.
- Ultraviolet probe (HD30.S2): 220 to 400 nm; calculates UVA, UVB and UVC irradiance; 2.5 nm bandpass; calibrated with a standard deuterium lamp.
- Detector: 2,048-element linear CCD, transmission diffraction-grating spectrometer; wavelength accuracy 0.3 nm and reproducibility 0.1 nm; stray light < 0.03%.
- Integration time: 1 ms to 4 s, automatic or manual.
- Stated uncertainties (visible probe): spectral irradiance ±5%; illuminance ±4%; PAR ±4%; CCT ±45 K; x,y coordinates ±0.002; CRI ±1.5. UV probe: spectral irradiance ±15%; UVA ±6%; UVB ±8%; UVC ±10%.
- Measurement modes: spectrum, photometry and colour, radiometry and spectral transmittance; single, continuous, monitor and automatic logging at 3, 5, 10, 15, 30 or 60-minute intervals.
- Data: 150 MB internal memory, micro SD card and USB stick; PC connection via Ethernet or mini USB; text export with the spectrum in 1 nm steps.
- Use: working temperature 0 to 40 °C; stated working range in illuminance of 5 to 70,000 lux.
Two of these numbers define where to use the instrument:
- 70,000 lux. Using Thimijan and Heins' conversion for sunlight (about 54 lux per µmol m⁻² s⁻¹), 70,000 lux is roughly 1,300 µmol m⁻² s⁻¹. Midday sun on a clear summer day exceeds that — Pinto et al. (2014) recorded 1,970 µmol m⁻² s⁻¹ in full sun in the UFLA experiment. Growth chambers, in vitro rooms, shaded greenhouses, forest understorey and environments under nets are within range; for full sun, plan measurements for times or conditions of lower irradiance and record them.
- 0 to 40 °C. The inside of a greenhouse at midday can exceed 40 °C. Do not leave the instrument logging continuously in that environment during the hottest hours.
The UV probe opens another front: UV-C germicidal lamps in laboratories, chambers with UV-B supplementation and material-ageing tests. For continuous outdoor UV-A at the weather station, the way to go is a dedicated radiometer, as discussed in ultraviolet radiation at the weather station.
Combining it with the weather station and continuous sensors
The spectroradiometer answers which light; continuous sensors answer how much and when. A robust design combines the two:
- A continuous quantum sensor in each environment (shelf, chamber, greenhouse bay), logging PPFD every few minutes. From it you calculate the daily light integral (DLI): DLI (mol m⁻² d⁻¹) = sum of PPFD × Δt (s) ÷ 1,000,000.
- A pyranometer or quantum sensor at the outdoor station, to relate the light inside the greenhouse to open-sky radiation and calculate the transmitted fraction over the cycle. Global radiation measurement is covered in the pyranometer at the weather station.
- Spectral campaigns at the beginning, middle and end of the experiment, using the same grid of points, to document the composition of the light and the drift of sources and materials.
A simple consistency check: under a broad-spectrum source (daylight or white LED), the PPFD calculated by the spectroradiometer and the one read by the quantum sensor, at the same point and time, should agree within the combined uncertainties of both instruments. If they do not, investigate levelling, shading, calibration or the quantum sensor's spectral response before going on with the experiment.
Common mistakes
- Describing the treatment by colour name ("blue LED") without measured peak, bandwidth and PPFD.
- Equalising treatments by lux or by the lamps' electrical power.
- Measuring a single point in the centre of the shelf and assuming it applies to the whole area.
- Ignoring the container in in vitro culture: the vessel lid and wall filter the light.
- Using catalogue data for nets and films instead of measuring on site and over time.
- Saturating the detector with too long an integration time and not noticing the clipped peak.
- Not stating the bands used for R:FR, or whether the reported PAR is 400–700 or 400–750 nm.
- Forgetting UV: in studies of secondary metabolites or with UV-blocking films, the 280–400 nm spectrum is also part of the treatment.
How to choose
- I need to characterise the spectrum of LEDs, nets, films and vessels, including UV: the HD30.1 spectroradiometer data logger, with visible and ultraviolet probes and a transmittance mode.
- I only need spot PPFD with a handheld meter: the LightScout quantum meters, 400 to 700 nm, with a 0 to 1,999 µmol m⁻² s⁻¹ display range.
- I need to log PPFD at two points and follow light alongside photosynthesis measurements: the ULM-500 universal light meter and data logger, with two PAR sensor inputs and memory for 50,000 lines.
- I need continuous wireless PPFD in several chambers or greenhouse bays: the Aranet wireless PAR light sensor, transmitting every 1, 2, 5 or 10 minutes.
- I need PAR at the outdoor weather station: the LightScout quantum PAR light sensor, compatible with WatchDog stations.
Related products
- HD30.1 spectroradiometer data logger: main instrument of this guide.
- LightScout quantum meters (ref. 3415Fxxx): spot PPFD from 400 to 700 nm.
- ULM-500 universal light meter and data logger (ref. ULM-500): two PAR channels with logging.
- Aranet wireless PAR light sensor (ref. TDSKAR01): continuous PPFD in greenhouses and chambers.
- LightScout quantum PAR light sensor (ref. 3668I): PAR for WatchDog stations.
- Radiometric probe for UVA irradiance (ref. LPUVA02): continuous outdoor UV-A, 315 to 400 nm.
- Spectrally flat Class B pyranometer (ref. LPPYRA02): global radiation at the outdoor station, 0.3 to 3 µm.
Checklist before using the spectrum in your thesis
- Do the methods report PPFD (400–700 nm) and, if far-red is present, ePAR (400–750 nm)?
- Were the blue, green, red and far-red fractions calculated from the measured spectrum?
- Are the R:FR bands stated?
- Was the measurement taken at the canopy plane or the top of the vessels, with the diffuser level?
- Is there a grid of points per shelf or environment, with mean and coefficient of variation?
- Was the transmittance of nets, films and containers measured, not just copied from the catalogue?
- Were spectral campaigns repeated throughout the experiment?
- Did a quantum sensor log PPFD continuously, and was the DLI calculated?
- Was the raw spectrum, in 1 nm steps, archived together with the calculation script?
If the answer is "yes" to all of them, the light in your experiment stops being a catalogue description and becomes a measured variable.
References
- McCree, K. J. The action spectrum, absorptance and quantum yield of photosynthesis in crop plants. Agricultural Meteorology, v. 9, p. 191–216, 1972.
- Zhen, S.; Bugbee, B. Far-red photons have equivalent efficiency to traditional photosynthetic photons: implications for redefining photosynthetically active radiation. Plant, Cell & Environment, v. 43, p. 1259–1272, 2020. DOI: 10.1111/pce.13730.
- Sager, J. C.; Smith, W. O.; Edwards, J. L.; Cyr, K. L. Photosynthetic efficiency and phytochrome photoequilibria determination using spectral data. Transactions of the ASAE, v. 31, n. 6, p. 1882–1889, 1988. DOI: 10.13031/2013.30952.
- Thimijan, R. W.; Heins, R. D. Photometric, radiometric, and quantum light units of measure: a review of procedures for interconversion. HortScience, v. 18, p. 818–822, 1983.
- Blonquist, M. Estimating phytochrome photoequilibrium with red/far red sensors and PAR-FAR sensors. Technical note, Apogee Instruments.
- Silva, R. A. L.; Soares, J. D. R.; Dias, G. M. G.; Pasqual, M.; Chagas, E. A.; Gavilanes, M. L. Cultivo de tamarindo sob malhas coloridas: plasticidade anatômica foliar. Ciência Rural, v. 45, p. 238, 2015. DOI: 10.1590/0103-8478cr20140012.
- Pinto, J. E. B. P.; Ferraz, E. O.; Bertolucci, S. K. V.; Silveira, H. R. O.; Santos, A. R.; Silva, G. M. Produção de biomassa e óleo essencial em mil folhas cultivada sob telas coloridas. Horticultura Brasileira, v. 32, p. 321, 2014. DOI: 10.1590/S0102-05362014000300014.
- Silva, M. M. A.; Oliveira, A. L. B.; Oliveira-Filho, R. A.; Camara, T.; Willadino, L.; Gouveia-Neto, A. The effect of spectral light quality on in vitro culture of sugarcane. Acta Scientiarum. Biological Sciences, v. 38, n. 2, p. 157, 2016. DOI: 10.4025/actascibiolsci.v38i2.31109.
- Santos Junior, C. F.; Rech, T. D.; Costa, M. D.; Navroski, M. C.; Boff, P.; Boff, M. I. C. Efeito da qualidade de luz na multiplicação e enraizamento in vitro de Cedrela odorata L. Ciência Florestal, v. 32, p. 2419, 2022. DOI: 10.5902/1980509866513.
- Delta OHM / Senseca. HD30.1 – Spectroradiometer / Data logger: instruction manual and datasheet.
- Hensistemas product sheet: HD30.1 spectroradiometer data logger.
Instruments for this application

DeltaOhm - GHM GroupHD30.1
HD30.1 Spectroradiometer Data Logger

Spectrum Technologies3415Fxxx
LightScout Quantum Meters

WalzULM-500
ULM-500 Universal Light Meter & Data Logger

AranetTDSKAR01
Aranet Wireless PAR Light Sensor

Spectrum Technologies3668I
LightScout Quantum PAR Light Sensor

DeltaOhm - GHM GroupLPUVA02
LPUVA02 Radiometric Probe for UVA Irradiance

DeltaOhm - GHM GroupLPPYRA02