Ultraviolet radiation at the weather station: measuring UVA with a radiometer for research

A teaching guide to UV measurement: UVA, UVB and UVC bands, why UVA irradiance is not the UV Index, how a UVA radiometer works, calibration, installation, dose and quality control.

By HenSistemas · · 8 min read

UVA radiometer with glass dome and white protective screen for ultraviolet measurement at a weather station

When people think of solar radiation at a weather station, the first instrument that comes to mind is the pyranometer, which measures energy across the whole solar spectrum. But a small fraction of that spectrum, the ultraviolet (UV), has outsized effects: it degrades polymers and coatings, drives atmospheric photochemistry, affects plants and microorganisms, and is the main cause of erythema (sunburn) and of much skin cancer. Measuring UV continuously, with a well-chosen and well-calibrated sensor, opens research lines a pyranometer alone cannot serve.

This guide is for faculty, graduate students and researchers at universities and public institutes who want to add UV to their station. We explain the UVA, UVB and UVC bands, the difference between UVA irradiance and the UV Index, how a solid-state UVA radiometer works, why UV calibration is more delicate than in the visible, and how to install, log and quality-control the data. For the broader radiation context, see our articles on the pyranometer and on diffuse, direct radiation and sunshine duration.

The ultraviolet bands

By the International Commission on Illumination (CIE) convention, UV is split into three bands:

  • UVC (100–280 nm): fully absorbed by ozone and oxygen in the upper atmosphere. It does not reach the surface.
  • UVB (280–315 nm): strongly absorbed by stratospheric ozone. Little reaches the ground, but it is biologically very active and dominates the erythemal effect.
  • UVA (315–400 nm): weakly absorbed by ozone. It is most of the UV reaching the ground, typically around 95% of surface UV energy at midday.

In energy terms UV is only a few percent of global radiation, so a pyranometer cannot be used to estimate it: the UV/global ratio varies with cloud, aerosols, ozone, solar zenith angle and altitude.

UVA irradiance is not the UV Index

This is the most common source of confusion in research projects.

  • UVA irradiance is a physical quantity: power per area (W/m²) in the 315–400 nm band, with no biological weighting. This is what a UVA radiometer measures.
  • The UV Index (UVI), defined by WHO, WMO, UNEP and ICNIRP, is irradiance weighted by the CIE erythemal action spectrum and multiplied by 40 m²/W. Because that action spectrum weights UVB hundreds to thousands of times more than UVA, the UVI depends mainly on UVB.

Practical consequence: a UVA sensor does not measure the UV Index. Empirical UVA–UVI correlations exist for a given site and season, but they shift with ozone and sun angle and do not replace an erythemal radiometer. If the research question is public health or erythemal dose, you need a different instrument. If it is materials degradation, photocatalysis, weathering tests, plant and insect photobiology (many insects see in the UVA) or characterising the site's radiation climate, UVA irradiance is exactly the quantity of interest.

In Brazil, INPE publishes UV Index forecasts, and summer values often reach the "very high" and "extreme" categories over much of the country. A local UVA series complements these forecasts with data measured at the experimental site itself.

How a solid-state UVA radiometer works

A weather-station UVA radiometer is essentially a photodiode with optical filters that restrict the response to UVA, under a diffuser and a dome. For the reference model in this guide, the Delta OHM outdoor UVA radiometer (LPUVA02), the manufacturer (now Senseca) states:

  • spectral response centred at 365 nm, half response between 342 and 384 nm and 1/100 between 320 and 400 nm (current datasheet);
  • measuring range 0 to 200 W/m², 2π sr field of view (hemisphere);
  • directional (cosine) response error below 8% from 0° to 80°;
  • response time below 0.5 s (95%), non-linearity below 1%, temperature coefficient below 0.1%/°C and long-term instability below ±3% per year;
  • operation from −40 to +80 °C, 4-pole M12 connector, 50 mm dome, silica-gel cartridge and bubble level.

The passive version needs no power and outputs a microvolt signal: typical sensitivity is 70 to 200 µV per W/m², and each unit's exact factor is given in its calibration report. Irradiance is simply E = V / S, where V is the measured voltage and S the calibration factor. Because the signal is small (the manufacturer typically quotes under 50 mV), the logger needs a low-voltage differential input with good resolution.

For a robust signal over long cables there are amplified versions: 4–20 mA output and voltage output (0–1, 0–5 or 0–10 V), powered at 10–30 Vdc (15–30 Vdc for 0–10 V). Confirm the output range of the supplied version when requesting a quote.

Why UV calibration is delicate

In the visible and infrared, pyranometers follow ISO 9060 and are traceable to world references. For UVA there is no equivalent international calibration standard for broadband radiometers, and the manufacturer itself warns that a calibration factor only makes sense together with the method used to obtain it. The LPUVA02 is individually calibrated against the primary standard of the manufacturer's laboratory, using the 365 nm line of a filtered Xe-Hg lamp.

The reason is physical: no filter reproduces a perfect 315–400 nm step. If the measured source spectrum differs from the calibration spectrum, spectral error can be significant. Larason and Cromer (NIST, 2001) discuss these error sources in detail. For research this means:

  1. Record the calibration method and factor in the series metadata.
  2. Do not compare absolute values between sensors from different manufacturers without a side-by-side intercomparison.
  3. Recalibrate periodically (the manufacturer recommends an annual check) and document factor changes.
  4. Where the absolute value matters for a paper, run an intercomparison against a reference spectroradiometer if one is available through a research network or institute.

Broadband UV sensor in µmol/m²·s

The catalogue also includes the LightScout UV light sensor (3676I) from Spectrum Technologies, which responds between 250 and 400 nm and works with WatchDog weather stations and micro-stations. It reports photon flux (0–200 µmol/m²·s, ±5%, 0–2 V output), the same unit used for PAR. It is a practical choice for groups already running WatchDog in greenhouses or the field who want an integrated UV indicator, but the unit and band differ from UVA irradiance in W/m². Do not mix the two series without an explicit spectral conversion.

Installation

  • Horizontal and level: use the bubble level. Levelling error adds to cosine error, which is already larger at high zenith angles.
  • Clear horizon: like a pyranometer, the UV radiometer must see the whole sky, with no obstacles above the sensor plane. UV has a large diffuse fraction, so obstacles matter even when they cast no direct shadow.
  • Height and access: mount on a mast bracket (the manufacturer offers one for 30–50 mm masts) where the dome can be cleaned often.
  • Away from bright reflective surfaces below the sensor, such as metal roofs or whitewashed walls.
  • Cable: use shielded cable and, for the passive version, the shortest possible run to the logger.

Logging and processing

A typical station routine for UV:

  1. Sample every 1–10 s and average over 1–10 minutes, aligned with the other radiation variables.
  2. Dose (radiant exposure): integrate irradiance over time. With 10-minute means in W/m², daily dose in kJ/m² is the sum of E × 600 s / 1000.
  3. UVA/global ratio: with a pyranometer on the same mast, the ratio helps identify thin cloud, biomass-burning aerosols and instrument problems.
  4. Night and offsets: the signal should be zero at night. A persistent night offset points to cable, grounding or logger problems.

Quality control

  • Negative values or values above the site's physical limit indicate failure.
  • Compare consecutive clear-sky days: the curve should be smooth and symmetric around solar noon.
  • Check the dome: dust, bird droppings and condensation reduce the signal. UV is more sensitive to soiling than the visible.
  • Replace the silica gel when it changes colour.
  • Log every cleaning, cable change and recalibration in the station diary.

Checklist before requesting a quote

  1. Is the research question about physical UVA (materials, photobiology, radiation climate) or about UV Index / erythemal dose? Only the first is served by a UVA sensor.
  2. Which output does your logger accept: mV, 4–20 mA or voltage?
  3. What cable length to the logger?
  4. Is power available at the mast for an amplified version?
  5. Will there be a pyranometer on the same mast for the UVA/global ratio?
  6. What is the recalibration and intercomparison plan?

References

  • CIE. Erythema reference action spectrum and standard erythema dose. ISO 17166:1999 / CIE S 007/E:1998.
  • Larason, T. C.; Cromer, C. L. Sources of error in UV radiation measurements. Journal of Research of the National Institute of Standards and Technology, v. 106, n. 4, 2001.
  • WHO; WMO; UNEP; ICNIRP. Global Solar UV Index: a practical guide. Geneva: WHO, 2002.
  • WMO. Guide to Instruments and Methods of Observation (WMO-No. 8), Volume I, radiation chapter.
  • Senseca (Delta OHM). LPUVA02 UVA radiometer datasheet and manual.