Dissolved oxygen in research: how to measure it with an optical sensor, correct for temperature, pressure and salinity, and combine the data with the weather station

A teaching guide to dissolved oxygen in research: concentration and saturation, Brazil's Conama 357 limits, the diel cycle and stratification, how the optical luminescence sensor works, temperature, barometric pressure and salinity corrections, calibration and Winkler checks, field installation, and estimating ecosystem metabolism with wind, pressure and radiation from the weather station.

By HenSistemas · · 15 min read

Submersible optical dissolved oxygen sensor with a black body, cable, 1-inch thread and protected tip

Dissolved oxygen (DO) is, arguably, the variable that best summarizes the health of a water body. It is produced by the photosynthesis of algae and macrophytes, enters through the surface from the atmosphere, and is consumed by the respiration of fish, plankton, bacteria, and by the decomposition of organic matter. When consumption outpaces production and exchange with the air, oxygen drops, and aquatic life feels it before any other indicator. That is why DO appears in water classification standards, water quality indices, aquaculture management, and in practically every limnology project.

This guide is for professors, graduate students, and researchers at universities and public institutes who need to measure DO with research-grade quality: in continuous monitoring of reservoirs, rivers, and ponds, or in field campaigns. The aim is to explain what the sensor measures, why temperature, atmospheric pressure, and salinity change the result, how to calibrate it, and why it is worth cross-referencing the data with a weather station installed next to the water.

What dissolved oxygen is and the units it is reported in

DO is the amount of oxygen gas (O₂) dissolved in water, expressed in two forms that should not be confused:

  • Concentration, in mg/L (equivalent to ppm in fresh water): the mass of O₂ available per liter. This is the unit used in standards and the one that matters to organism biology.
  • Saturation, in %: the ratio between the measured concentration and the concentration the water would have in equilibrium with air under the same conditions of temperature, pressure, and salinity. Values above 100% (supersaturation) indicate intense photosynthetic production or aeration; values well below 100% indicate consumption.

The relationship between the two depends on the solubility of oxygen, which varies considerably with conditions. In fresh water, at sea level (760 mmHg), saturated water contains about 9.09 mg/L at 20 °C, 8.26 mg/L at 25 °C, and 7.56 mg/L at 30 °C, according to the solubility tables used by the USGS (DOTABLES program) and by ISO 5814. In water with a salinity of 36 ‰, at 25 °C, solubility drops to about 6.7 mg/L. In other words: the same reading of 7 mg/L means well-oxygenated water in a tropical pond at 30 °C, but water already showing a deficit in a cold mountain stream.

How much oxygen is "enough": the Conama 357 reference

In Brazil, Conama Resolution No. 357/2005 sets minimum DO limits for freshwater classes: not less than 6 mg/L in Class 1, 5 mg/L in Class 2, 4 mg/L in Class 3, and above 2 mg/L in Class 4. DO is also one of the nine variables in the Water Quality Index (IQA) used by Cetesb and by the National Water and Basic Sanitation Agency (ANA), carrying the highest weight among them (0.17), expressed as percent saturation.

For research purposes, these references show two things. First, that measurement needs accuracy compatible with limits that are only 1 mg/L apart. Second, that a single sample collected at 10 a.m. may be far from the day's minimum, as shown below.

The daily cycle: why a single reading is misleading

In waters with primary production, DO follows a daily cycle driven by light. During the day, photosynthesis adds oxygen and concentration rises, often reaching supersaturation in the mid-afternoon. At night, photosynthesis stops, respiration continues, and DO falls, reaching its minimum near dawn. In aquaculture ponds and eutrophic reservoirs, this amplitude can span several mg/L, and it is in the early hours before dawn that fish kills occur.

Hence the first practical lesson: to characterize an environment, measure continuously. A fixed sensor logging every 10 to 15 minutes reveals the amplitude, the timing of the minimum, and the response to cold fronts, cloudy days, or strong winds. Spot campaigns remain useful for mapping spatial variation and vertical profiles, but always noting the time and weather conditions.

In deep water bodies, there is also thermal stratification: in summer, the warmed surface layer (epilimnion) separates from the bottom layer (hypolimnion), which does not exchange gas with the atmosphere and can become oxygen-depleted. Profile measurement, with a probe lowered through several levels or with fixed sensors at different depths, is what reveals this pattern.

How the optical luminescence sensor works

For decades, DO in the field was measured with Clark-type electrochemical sensors (polarographic or galvanic): oxygen crosses a membrane and is reduced at a cathode, generating a current proportional to its partial pressure. These work well, but they consume oxygen during measurement (which is why they require water flow over the membrane), have an electrolyte that depletes, and drift over time.

The optical sensor uses a different principle. At the tip there is a cap with an immobilized luminescent dye. An LED excites the dye, which emits light back. Oxygen molecules, upon colliding with the dye, "quench" part of this luminescence (a phenomenon called quenching): the more oxygen present, the lower the intensity and the shorter the lifetime of the emission. This relationship is described by the Stern–Volmer equation:

τ₀ / τ = 1 + K_SV · [O₂]

where τ₀ is the luminescence lifetime with no oxygen present, τ is the measured lifetime, and K_SV is the dye constant. In practice, the sensor measures the phase shift between the excitation light and the emitted light, which is proportional to τ and less sensitive to dye aging than intensity is.

The consequences for the researcher:

  • It does not consume oxygen, so it does not require water flow over the tip, an advantage in still waters, wells, and sediments.
  • It has no electrolyte or membrane to replace frequently; maintenance focuses on cleaning and periodically replacing the sensor cap.
  • It drifts less, which reduces the number of calibration visits needed at remote installations.
  • It is recognized in standardized methods, such as Standard Methods 4500-O H (optical probe) and ASTM D888 (luminescence method).

The point to keep in mind is that the sensor essentially measures oxygen's partial pressure. To convert this to mg/L, it needs to know the temperature, salinity, and, for saturation, the atmospheric pressure.

Temperature, pressure, and salinity: the three corrections

Temperature. This affects both the dye's response and solubility. Every DO sensor has a thermometer at the tip and performs automatic compensation; what the researcher should verify is whether this thermometer agrees with a reference, because a 1 °C error produces roughly a 2% error in the calculated saturation.

Atmospheric pressure. Solubility is proportional to the partial pressure of oxygen in the air, which depends on barometric pressure. The approximate correction is:

Cs(P) = Cs(760) · (P − p_w) / (760 − p_w)

where P is the local pressure and p_w is the vapor pressure of water at the sample temperature (both in the same unit). A numerical example: in a reservoir at about 800 m of altitude, with a pressure of 920 hPa and water at 25 °C (p_w ≈ 31.7 hPa), the factor is (920 − 31.7)/(1013.25 − 31.7) ≈ 0.905, and solubility drops from 8.26 to about 7.48 mg/L. A reading of 7.0 mg/L then corresponds to 94% saturation; if the calculation used sea-level pressure instead, the result would be 85%, a nine-point error. This is why calibration in saturated air and the saturation calculation must use the actual barometric pressure at the time, not a fixed value. A weather station next to the measurement point solves this; see our guide on atmospheric pressure measurement at weather stations.

Salinity. Dissolved salts reduce solubility. In fresh water the effect is small, but in estuaries, coastal lagoons, mariculture, and saline effluents it is decisive. Salinity can be entered as a fixed value or, better, calculated from a conductivity sensor installed alongside, as multiparameter probes do.

Calibration and verification

Routine calibration of an optical sensor involves two points:

  1. 100% saturation: the sensor is placed in air saturated with water vapor (for example, inside a calibration cup with a moist sponge, with no droplets on the cap), at a stable temperature, with the current barometric pressure entered into the instrument.
  2. Zero: in a sodium sulfite solution, which consumes all the oxygen. Many manufacturers recommend performing the zero check only periodically, since the optical sensor is stable at this point.

For research work, it is worth including an independent verification using the Winkler method (iodometric titration, Standard Methods 4500-O C), on samples collected alongside the sensor at the beginning and end of each deployment period. The difference before and after cleaning separates the effect of biofouling from sensor drift, a practice described in the DO chapter of the USGS National Field Manual (Rounds, Wilde, and Ritz, 2013).

Field installation

  • Depth and position: define the layer that answers the research question (surface, mid-column, near-bottom) and keep the sensor at the same depth consistently, using a float or fixed mount.
  • Protection: use a perforated protective tube that allows water to circulate but shields against branches and animals. In ponds, bite-resistant guards make a difference.
  • Biofouling: the main cause of poor data in long-term deployments. Plan regular cleanings, more frequently in summer and in eutrophic water; probes with a wiper or automatic cleaning help.
  • Logging: a 10- to 15-minute interval is sufficient for the daily cycle; also log water temperature and, if possible, conductivity.
  • Power and communication: at remote sites, prefer low-power sensors with a digital output (Modbus RTU or SDI-12) and a data logger with battery, solar panel, and telemetry, to track the data without visiting the site. Our guide on weather station telemetry discusses the options.

Cross-referencing DO with the weather station: ecosystem metabolism

With a continuous series of DO and weather data, it is possible to estimate the metabolism of a water body using the diel oxygen curve method, proposed by Odum (1956) and revised by Staehr and colleagues (2010). The DO change over an interval is decomposed into three terms:

ΔO₂ / Δt = GPP − R − F / z_mix

where GPP is gross primary production, R is ecosystem respiration, F is the exchange of oxygen with the atmosphere, and z_mix is the depth of the mixing layer. At night, GPP = 0, and the variation gives respiration directly; during the day, adding respiration back, production can be obtained.

The air-exchange term is F = k (O₂ − O₂sat), and this is where the weather station comes into play:

  • Wind at 10 m (U₁₀) controls the transfer coefficient k. A widely used model for lakes is k₆₀₀ = 2.07 + 0.215 × U₁₀ raised to the power of 1.7 (k₆₀₀ in cm/h and U₁₀ in m/s), attributed to Cole and Caraco (1998), which is then converted for oxygen using the Schmidt number. See our guide on wind measurement.
  • Atmospheric pressure and water temperature define O₂sat at each interval.
  • Solar radiation or photosynthetically active radiation (PAR) marks the start and end of the photosynthesis period and allows production to be modeled as a function of light.

This cross-referencing turns a water quality sensor into an ecosystem ecology instrument, useful in studies of eutrophication, the effects of rainfall and cold fronts, the impact of sewage loads, and pond management.

Processing and quality control

  • Remove maintenance periods and the first few minutes after each reinstallation.
  • Compare readings with Winkler verifications or with a reference sensor and apply a linear drift correction between visits, when justified.
  • Flag values outside the physically plausible range (negative values or values well above plausible supersaturation) and abrupt jumps with no meteorological explanation.
  • Recalculate saturation using the measured barometric pressure, not a fixed value.
  • Keep metadata: sensor depth, cleaning and calibration dates, cap replacements, and the salinity value used in compensation.

Common mistakes

  1. Calibrating in saturated air while entering sea-level pressure at a high-altitude site.
  2. Forgetting the salinity correction in brackish water.
  3. Comparing spot readings taken at different times of day.
  4. Leaving the sensor uncleaned for weeks in eutrophic water.
  5. Installing the sensor in a dead zone, resting on the bottom or inside a tube with no circulation.
  6. Not logging water temperature alongside DO.

The catalog's optical sensor

As an example of a fixed sensor for continuous monitoring, the catalog offers the optical dissolved oxygen sensor with 4-20 mA and RS485 outputs (ref. ODOS-5000). According to the product datasheet:

  • measurement range of 0 to 20 mg/L and 0 to 200% saturation, with wider ranges available on request;
  • accuracy of ±1% at 25 °C and response time (t90) under 60 s;
  • automatic temperature compensation, operating range of −5 to +50 °C, and maximum pressure of 2 bar;
  • suitable for fresh and salt water, with no oxidizable parts, and requiring no flow over the tip;
  • 4-20 mA (2-wire) and RS485 Modbus RTU outputs, plus an interface that emulates a Clark-type sensor, for replacing electrochemical sensors in existing systems; for research use, it can be connected directly to a computer via an optional USB converter;
  • replaceable sensor cap, automatic cap-fault diagnostics, a PVC body 20 mm in diameter (30 mm in the submersible version), and a submersible version with bite protection for fish farming, rated to 50 m depth.

For researchers, the Modbus RTU output simplifies integration with weather station data loggers, and the fact that it requires no flow allows installation in ponds and still-water lakes. Two points to consider when designing a project: the datasheet does not mention automatic correction for barometric pressure or salinity, which can be handled in post-processing using data from the weather station and a conductivity sensor; and, as with any long-term deployment, biofouling requires a cleaning routine.

Catalog alternatives and complements

  • Dissolved oxygen and temperature measurement and control system (ref. ODOS-3000): a controller that works with the optical sensor, displaying DO in %, mg/L, or ppm and temperature, with two relay outputs and one 4-20 mA output, accepting sensor cables up to 1000 m long, and rated IP67. Useful for triggering aerators in aquaculture experiments.
  • Aqua TROLL 400 multiparameter probe (In-Situ): optical DO with accuracy of ±0.1 mg/L from 0 to 8 mg/L, plus conductivity, pH, ORP, temperature, and level; automatic salinity compensation and, with the manufacturer's system, barometric pressure compensation; Modbus/RS485 and SDI-12 outputs. Suited for long-term fixed stations.
  • Multiparameter probe with up to 12 sensors (ref. MANTA PLUS, Eureka): optical DO, pH, conductivity, turbidity, and fluorometers such as chlorophyll a and phycocyanin, with internal logging; larger configurations include a central wiper cleaning system. Well suited for eutrophication and cyanobacteria bloom studies.
  • Portable multiparameter probe (ref. AP-5000, Aquaread): optical DO, conductivity, pH, ORP, salinity, temperature, and depth, with cables from 3 to 30 m. For field campaigns and vertical profiling.
  • Conductivity and temperature sensor with SDI-12 and 4-20 mA outputs (ref. EC1550): a low-power toroidal sensor that provides the conductivity used to calculate the salinity applied in DO compensation.
  • Data logger with telemetry for SDI-12 sensors (ref. HS14GS-SDI12): solar powered, IP68 enclosure, and 3G/4G module, for bringing data from SDI-12 probes at a remote site back to the lab.

For further reading, see also our guides on what to monitor in water quality and on water level measurement with a pressure transducer and barometric compensation.

Checklist before budgeting

  1. What is the research question: classification compliance, metabolism, pond management, effluent impact, or stratification?
  2. Will the measurement be continuous (fixed sensor) or campaign-based (portable probe), and at how many depths?
  3. What accuracy is required near the Conama 357 limits?
  4. Is the water fresh, brackish, or saline, and will there be a conductivity sensor for compensation?
  5. Is there a weather station nearby, providing pressure, wind, and radiation, to calculate saturation and metabolism?
  6. What data logger and protocol (Modbus RTU, SDI-12, 4-20 mA), at what interval and with what telemetry?
  7. What will the routine be for cleaning, calibration, and Winkler verification?

References

  • Brasil. Conselho Nacional do Meio Ambiente. Resolução Conama nº 357, de 17 de março de 2005 (classificação dos corpos de água e padrões de qualidade).
  • Agência Nacional de Águas e Saneamento Básico (ANA). Portal da Qualidade das Águas: Índice de Qualidade das Águas (IQA).
  • U.S. Geological Survey. DOTABLES: dissolved oxygen solubility tables (com base em Benson e Krause, 1980, 1984).
  • ISO 5814:2012. Water quality – Determination of dissolved oxygen – Electrochemical probe method (tabelas de solubilidade).
  • Rounds, S. A.; Wilde, F. D.; Ritz, G. F. Dissolved oxygen. In: National Field Manual for the Collection of Water-Quality Data, Book 9, Chapter A6.2. U.S. Geological Survey, 2013.
  • APHA; AWWA; WEF. Standard Methods for the Examination of Water and Wastewater, métodos 4500-O C (Winkler, modificação azida) e 4500-O H (sonda óptica).
  • ASTM D888. Standard Test Methods for Dissolved Oxygen in Water (método C, luminescência).
  • Odum, H. T. Primary production in flowing waters. Limnology and Oceanography, v. 1, n. 2, p. 102–117, 1956.
  • Staehr, P. A.; Bade, D.; Van de Bogert, M. C.; Koch, G. R.; Williamson, C.; Hanson, P.; Cole, J. J.; Kratz, T. Lake metabolism and the diel oxygen technique: state of the science. Limnology and Oceanography: Methods, v. 8, p. 628–644, 2010.
  • Cole, J. J.; Caraco, N. F. Atmospheric exchange of carbon dioxide in a low-wind oligotrophic lake measured by the addition of SF6. Limnology and Oceanography, v. 43, n. 4, p. 647–656, 1998.