Meteorology · Water · Technical guides

Groundwater and surface water level in research: measuring with a pressure transducer and correcting for barometric pressure

A teaching guide to water level measurement in wells and rivers: hydrostatic pressure, absolute vs vented sensors, barometric compensation, barometric efficiency, range selection, installation, recharge and quality control.

By HenSistemas · · 13 min read

Cylindrical water level logger with dark coating and suspension eyelet, for installation in a monitoring well

In hydrology and hydrogeology, water level is the simplest variable to state and one of the easiest to measure badly. An observation well or a river section logged every 15 minutes for years reveals recharge, pumping drawdown, the response to intense rainfall, tidal effects and even the aquifer "breathing" with atmospheric pressure. But if barometric compensation, the elevation reference and manual verification are neglected, the series can carry errors of several centimetres that no later statistics will fix.

This guide is for lecturers, graduate students and researchers at public universities and institutes who will install level loggers in wells, piezometers, streams, ponds or reservoirs. We explain the physics of the pressure transducer, the difference between absolute and vented sensors, how to do barometric compensation, how to choose the range, how to install and reference the logger, how to use the data (recharge, aquifer tests, rating curves) and how to control quality. The link to the weather station is direct: the site barometer and rain gauge are part of the same analysis. For context, see also our articles on atmospheric pressure at the weather station and on the tipping bucket rain gauge.

The principle: hydrostatic pressure

The level logger most used in research is a submerged pressure transducer. It measures the pressure on its diaphragm, and the level follows from the hydrostatic law:

P = P_atm + ρ · g · h

where P is the measured pressure, P_atm the atmospheric pressure on the free surface, ρ the water density, g the acceleration of gravity and h the height of the water column above the sensor. Solving for h:

h = (P − P_atm) / (ρ · g)

Useful orders of magnitude for the classroom and the field:

  • With ρ ≈ 997 kg/m³ (fresh water at 25 °C) and g = 9.80665 m/s², 1 kPa corresponds to about 10.2 cm of water column, i.e. 1 hPa ≈ 1 cm.
  • Sea-level atmospheric pressure (101.3 kPa) equals a little over 10 m of water column. An absolute sensor "sees" those 10 m all the time, even out of the water.
  • Atmospheric pressure varies with the daily barometric tide (well marked in the tropics) and with frontal systems. During cold fronts in southern and south-eastern Brazil, changes of about 10 hPa within a few days are common, which equals roughly 10 cm of apparent level if uncompensated.

That is why, in an unconfined-aquifer well with a true change of a few centimetres per week, the uncompensated series mostly shows the weather, not the hydrology.

Absolute or vented sensor?

There are two physical ways to remove P_atm from the equation:

Absolute (non-vented) Vented (gauge)
What it measures Water + atmosphere Water only (the vent tube brings P_atm to the back of the diaphragm)
Compensation Requires an on-site barometer and processing Done physically by the cable
Cable Can be a simple wireline or cord Continuous vented cable to the surface
Weak points Depends on barometer quality and synchronisation Moisture in the vent tube, desiccant to replace, costlier cable more prone to damage
Typical use Well networks, remote sites, discreet locked installation Continuous telemetry with a surface logger

Brazil's Integrated Groundwater Monitoring Network (RIMAS), planned by the Geological Survey of Brazil (SGB) from 2009, now has more than 500 wells in 22 states. According to SGB, the network uses pressure-sensor loggers, both with vented cable and with a second logger dedicated to atmospheric pressure, as well as float-and-counterweight recorders, and checks the records against manual readings at each operating visit. It is a good procedural reference for academic projects.

Barometric compensation in practice

With an absolute sensor, the compensated column is simply:

h(t) = [P(t) − P_baro(t)] / (ρ · g)

In practice, four precautions determine quality:

  1. Same instant. The barometer must log at the same times as the submerged logger (or be interpolated to them). Use the same time zone, preferably UTC, on both instruments.
  2. Roughly the same place. The manufacturer of the reference logger in this guide recommends one dedicated barometer per area of about 30 km radius and per 300 m of elevation difference, installed above the maximum water level, preferably inside the well itself, where temperature is more stable.
  3. Same unit and density. Convert both series to the same unit before subtracting and record the ρ used. In brackish or sea water (ρ ≈ 1025 kg/m³), using fresh-water density overestimates the column by about 2.5%.
  4. Barometer accuracy matters. The error of the compensated level includes the barometer error. A barometer with ±0.05 kPa contributes about ±0.5 cm; one with ±1.5 hPa contributes about ±1.5 cm. The constant part of this error is removed when the series is referenced to a manual reading, but drift and hysteresis remain.

You can use the barometer of a nearby weather station, provided it is on the same site, with a synchronised clock and known accuracy. That is a good argument for integrating the station and the well network in the same project.

When the aquifer "breathes": barometric efficiency

In confined aquifers, an increase in atmospheric pressure is carried partly by the water and partly by the aquifer matrix. In the open well, the level falls when pressure rises. This effect is described by the barometric efficiency (BE), defined by Jacob (1940):

BE = − Δh_well / (ΔP_atm / ρg)

with BE between 0 (shallow unconfined aquifer, no effect) and 1. An absolute sensor in the well measures ΔP = ΔP_atm · (1 − BE). Teaching points:

  • In an unconfined aquifer, simple compensation is enough.
  • In a confined aquifer, the compensated series still shows an inverted barometric signature. It is not instrument error: it is aquifer physics. For pumping tests it must be removed; for aquifer characterisation it is information.
  • In unconfined aquifers with a thick unsaturated zone, the response can lag in time. Rasmussen and Crawford (1997) show how to identify and remove these effects by lagged regression.
  • In confined aquifers, earth tides also appear as semidiurnal oscillations of a few millimetres to centimetres.

Archiving the barometric series alongside the level series lets you compute BE, which is as valuable for theses as the level itself.

How to choose the logger

  • Nominal range. Accuracy is given as a percentage of full scale (% FS). Choose the smallest range that covers the maximum expected column above the sensor, with margin for floods and pumping drawdown. For the reference logger, accuracy is ±0.05% FS, giving ±0.3 cm on the 5 m range, ±0.5 cm on 10 m, ±1 cm on 20 m and ±1.5 cm on 30 m. An oversized range multiplies the error.
  • Temperature. The sensor's thermal compensation has a defined range. Water temperature, logged together, is also research data (river–aquifer interaction, recharge) and an installation diagnostic.
  • Memory and battery. Compute autonomy for the planned interval and visit frequency.
  • Integration. Stand-alone logger, SDI-12 transmitter connected to the station logger, or wireless sensor with an app. The choice depends on logistics and on whether real-time data are needed.

The reference logger in this guide

The main instrument in this guide is the Solinst stand-alone water level and temperature logger (Levelogger Edge, model 3001). From the product sheet and the manufacturer's datasheet:

  • absolute piezoresistive silicon pressure sensor with a Hastelloy diaphragm and automatic temperature compensation between 0 and 50 °C;
  • accuracy of ±0.05% FS, in ranges of 5, 10, 20, 30, 100 and 200 m of water column;
  • platinum temperature sensor (RTD), ±0.05 °C accuracy and 0.003 °C resolution;
  • memory for 40,000 sets of level and temperature readings, or up to 120,000 with compression in linear mode;
  • lithium battery with a 10-year life (based on one reading per minute) and a clock accurate to ±1 minute per year;
  • sampling from 1/8 s to 99 h, in linear, event-based or user-defined schedule modes;
  • 22 mm × 159 mm stainless steel body with titanium-based PVD coating, optical infrared communication and deployment on wireline, Kevlar cord or direct read cable;
  • the manufacturer's software barometrically compensates several files at once, with adjustments for density, elevation and barometric efficiency.

Being absolute, it needs an on-site barometer, as described above. In exchange, it fits a 2-inch well, can be locked out of sight and needs no vented cable.

Alternatives and complements from the catalogue

  • Level and temperature transmitter with SDI-12 output (HS1412L-SD12): silicon sensor with vented cable, range up to 200 m of water column, accuracy of ±0.25% FS or ±0.1% FS depending on version, PT100 for temperature, under 100 µA in sleep. This is the option for wiring level into the station logger with telemetry.
  • 4G data logger for weather stations (HD33MT.4): SDI-12 master port, analogue and counter inputs and mobile-network data transmission. It brings level, barometer and rain gauge together in one instrument.
  • Air pressure transmitter with SDI-12 output (SP1A): piezoresistive barometer with ±1.5 mbar accuracy, useful as the site barometer. Remember that ±1.5 hPa equals about ±1.5 cm of column.
  • Capacitance water level logger (Xtreem): measures level along a sensor cable, with ±1.0 mm accuracy and 0.15 mm resolution, independent of atmospheric pressure; data collected over Bluetooth with an Android app. Suited to shallow water, tanks, wetlands and short piezometers, within the length of the sensor cable.
  • Manual water level meter with laser-marked tape (HS14LM): stainless steel probe with audible and visual alert, 1 mm resolution. It is the reference instrument to install and check any automatic logger.

Installation in a well

  1. Define the reference point (usually the top of casing) and survey its elevation by levelling or GNSS. Mark it physically.
  2. Measure the level manually with the tape meter before lowering the logger: depth to water from the reference (DTW).
  3. Position the sensor below the lowest expected level, including drawdown, but above the bottom so that sediment does not bury it. Use low-stretch wireline or Kevlar and record the exact length.
  4. Wait for thermal equilibrium for a few minutes before taking the reference reading.
  5. Vented cap: the well must be in equilibrium with the atmosphere; an airtight cap creates its own pressure inside the well.
  6. Barometer above the maximum level, synchronised with the submerged logger.

Conversion to depth and elevation uses the initial manual reading:

DTW(t) = DTW_ref − [h(t) − h_ref] Water level elevation(t) = Reference elevation − DTW(t)

where h_ref is the compensated column at the time of the manual reading DTW_ref.

Installation in a river, stream or pond

  • Install the sensor inside a perforated PVC stilling well, anchored to a firm structure, protected from current and from debris impact during floods.
  • Install a staff gauge and read it at every visit: it is the level reference, just as the manual reading is in the well.
  • Watch for sedimentation, biofouling and sensor exposure in the dry season.
  • To turn stage into discharge you need the rating curve, Q = a · (h − h₀)^b, fitted to discharge measurements at different stages. Continuous stage only becomes continuous discharge with those field measurements.

Using the data in research

  • Recharge by the water-table fluctuation (WTF) method. In an unconfined aquifer, R ≈ S_y · Δh / Δt, where S_y is the specific yield and Δh the rise after an event, measured against the extrapolated recession curve (Healy and Cook, 2002). The site rain gauge is essential to attribute each rise to a rainfall event. Wendland, Barreto and Gomes (2007), at the University of São Paulo in São Carlos, used well monitoring, a rain gauge and a stage recorder in the Ribeirão da Onça basin, an outcrop zone of the Guarani Aquifer System, to study recharge and the water balance.
  • Pumping and slug tests. They need fast sampling at the start (fractions of a second to a few seconds) and sparser sampling later. Classic methods, such as Hvorslev (1951) for slug tests, rely on a good recovery series.
  • River–aquifer interaction and wetlands. Wells near a stream, with level and temperature, show whether a reach gains or loses water.
  • Irrigated agriculture. A shallow water table in floodplains governs drainage, root aeration and capillary contribution.

Quality control

  • Manual check at every visit. Compute the residual between logger level and manual reading. A trend in the residual indicates sensor drift, cable stretch or logger movement.
  • Residual barometric signature. In an unconfined aquifer, the compensated series should not correlate with pressure. If it does, check clock synchronisation, units and distance to the barometer; in a confined aquifer, evaluate BE.
  • Sensor out of water. A compensated column near zero or negative indicates a dry well or a sensor above the water level.
  • Temperature as a diagnostic. Abrupt jumps in water temperature indicate that the logger was pulled, moved, or that there was nearby pumping.
  • Interferences. Log neighbouring pumping, works and extreme events in the well diary.
  • Aggregation. RIMAS, for example, publishes the daily median of hourly levels; the median is robust to isolated spikes.
  • Metadata. Reference elevation, cable length, serial number, sensor range, ρ adopted, barometer used, time zone and check history.

Checklist before requesting a quote

  1. Well or surface water? What is the maximum column expected above the sensor, including floods and drawdown?
  2. Absolute or vented? Is there a barometer on site or at the nearby weather station?
  3. Must data arrive in real time (SDI-12 and telemetry), or is collection during visits enough?
  4. What logging interval and visit frequency? Do memory and battery cover that period?
  5. Is the water fresh, brackish or saline? Which ρ will be used?
  6. Is there a manual meter and a surveyed reference point for checks?
  7. Will there be pumping or slug tests requiring fast sampling?

References

  • Healy, R. W.; Cook, P. G. Using groundwater levels to estimate recharge. Hydrogeology Journal, v. 10, p. 91–109, 2002.
  • Hvorslev, M. J. Time lag and soil permeability in ground-water observations. Bulletin 36. Vicksburg: U.S. Army Corps of Engineers, Waterways Experiment Station, 1951.
  • Jacob, C. E. On the flow of water in an elastic artesian aquifer. Transactions, American Geophysical Union, v. 21, p. 574–586, 1940.
  • Rasmussen, T. C.; Crawford, L. A. Identifying and removing barometric pressure effects in confined and unconfined aquifers. Ground Water, v. 35, n. 3, p. 502–511, 1997.
  • Geological Survey of Brazil (SGB). Integrated Groundwater Monitoring Network (RIMAS).
  • Wendland, E.; Barreto, C.; Gomes, L. H. Water balance in the Guarani Aquifer outcrop zone based on hydrogeologic monitoring. Journal of Hydrology, v. 342, p. 261–269, 2007.
  • Solinst. Levelogger Edge, model 3001: datasheet and user guide.