Wet and dry atmospheric deposition in research: how to sample with an automatic collector, preserve the sample and combine rain chemistry with the weather station
A teaching guide to atmospheric deposition in research: the difference between wet, dry and bulk deposition, what rain chemistry reveals in Brazil, volume-weighted means and deposition fluxes, how the automatic wet/dry sampler works, selection and siting criteria, sample collection and preservation, ion and conductivity balances, and combining the data with rainfall and wind from the weather station.

Rain brings more than just water. Every drop that falls washes the atmosphere and carries to the soil, to rivers, and to leaves a load of ions, nutrients, acids, and metals that were in the air. On dry days, dust, ash, and aerosols settle onto surfaces. Together, these two pathways form atmospheric deposition, and measuring it is the way to answer questions that arise in ecology, agronomy, environmental chemistry, hydrology, and public health: is the rain in my region acidic? How much nitrogen does the forest or the reservoir receive from the air each year? Do sugarcane burning, traffic, or sea spray change the composition of the rain?
This guide is for professors, graduate students, and researchers at universities and public research institutes who are setting up a deposition sampling site. The goal is to explain what separates wet, dry, and total deposition, how the automatic wet/dry sampler (the wet & dry deposition gauge) works, how to install and operate it without contaminating the sample, how to validate laboratory results, and why the data is only complete when cross-referenced with a weather station.
Wet, dry, and total deposition: three different things
Before purchasing any equipment, it is worth fixing the terms, because each corresponds to a different collection method:
- Wet deposition: what reaches the ground dissolved or suspended in rain, snow, hail, or drizzle. It is collected with a container that remains open only during precipitation (wet-only collection).
- Dry deposition: the transfer of gases and particles from the air to surfaces without the involvement of precipitation. In field practice, what an open container can actually capture is mostly the sedimentable fraction of particles (dust, ash, pollen). Gases such as SO₂, NO₂, and NH₃, and fine particles, deposit in a way that depends heavily on the surface (leaf, soil, water) and are normally estimated using inferential methods (air concentration × deposition velocity), not a bucket.
- Total (bulk) deposition: what falls into a funnel or bucket that is always open. It includes rain plus everything that has settled between one event and another. It is inexpensive and useful for nutrient balances in watersheds, but it mixes the two pathways and tends to overestimate the "rain" fraction.
The automatic wet/dry sampler solves the main limitation of the bulk collector: a precipitation sensor drives a movable lid that covers the wet collector when it is not raining and switches to cover the dry collector once rain begins. This way, the two fractions remain separated in distinct containers, and each one goes to the laboratory with its own history.
Why measure it: what rainwater chemistry reveals
Pure water in equilibrium with atmospheric CO₂ has a pH close to 5.6, and this is the threshold used in the literature to characterize acid rain. Below that value, there is a contribution from strong acids (sulfuric and nitric, formed from SO₂ and NOₓ) or from weak organic acids (formic and acetic). Above it, neutralization occurs via ammonia and carbonate dust.
Some Brazilian results show the kind of information that comes out of a well-run program:
- In downtown São Paulo, in wet-only collections between March 2002 and February 2003 (Universidade Presbiteriana Mackenzie, with participation from IAG-USP, the Institute of Astronomy, Geophysics and Atmospheric Sciences at the University of São Paulo), the rain had an average pH of 4.99 and a volume-weighted mean of 5.19; the most abundant ion was ammonium (37.6 µmol/L), followed by nitrate (21.2 µmol/L) and sulfate (12.4 µmol/L), with organic acids accounting for about 35% of the potential free acidity (Leal et al., 2004).
- In the Piracicaba River basin (state of São Paulo), 272 samples collected between 1997 and 1998 at four sites showed acid rain at all of them (average pH of 4.4 to 4.5), with composition controlled mainly by three sources: soil dust, sugarcane burning, and industrial emissions (Lara et al., 2001).
- Long time series from the São Paulo Metropolitan Region allowed analysis of trends in precipitation chemistry between 1983 and 2003, showing how changes in fuels and emissions show up in the rain (Fornaro and Gutz, 2006).
For the researcher, the concrete question is usually one of these: how much nitrogen and sulfur enter the ecosystem via the atmospheric pathway; what is the marine contribution (sodium, chloride, magnesium) in coastal areas; what is the effect of wildfires/burning during the dry season; how much metal reaches a water source; or how the dust from an industrial source is distributed in the surrounding area.
Units, weighted average, and deposition flux
Three calculations show up in every deposition study, and it is worth mastering them before planning the collection.
1. Concentration in equivalents. Laboratories report ions in mg/L or µmol/L. For charge balance, convert to µeq/L by multiplying the molar concentration by the charge: 12.4 µmol/L of sulfate (SO₄²⁻) corresponds to 24.8 µeq/L. H⁺ comes from pH: pH 5.0 is equivalent to 10 µmol/L of H⁺.
2. Volume-weighted mean (VWM). Light rains tend to be very concentrated, and heavy rains, diluted. The arithmetic mean of the samples overstates the weight of small rain events. For this reason, the following is used:
VWM = Σ (Cᵢ × Vᵢ) / Σ Vᵢ
where Cᵢ is the concentration and Vᵢ is the volume (or rainfall depth) of each sample. The VWM of pH is calculated from the H⁺ concentration, and only at the end is it converted back to the pH scale.
3. Wet deposition flux. The VWM is multiplied by the rainfall depth for the period. With concentration in mg/L and rainfall in mm:
Deposition (kg/ha) = C (mg/L) × P (mm) / 100
The factor of 100 comes from the fact that 1 mm of rain over 1 hectare equals 10,000 liters. An exercise using the São Paulo numbers cited above: ammonium + nitrate add up to 58.8 µmol/L of inorganic nitrogen, or about 0.82 mg N/L. For a hypothetical annual rainfall of 1,400 mm, that would amount to roughly 11.5 kg of N per hectare per year just in the dissolved inorganic fraction. This is an illustrative calculation, but it shows why atmospheric deposition enters the nutrient balances of forests, pastures, and reservoirs.
Note that the third calculation depends on a reliable rainfall depth, which comes from the rain gauge, not from the sampler. We will return to this point.
How the automatic wet/dry sampler works
An automatic deposition gauge has four parts:
- Precipitation sensor. It is generally a grid of electrodes over an insulating plate (conductivity detection) or a capacitive element. The raindrop closes the circuit or changes the capacitance, and the controller triggers the lid. The sensor is heated to a temperature slightly above ambient to dry quickly after rain (allowing the lid to close), to melt snow or hail, and to avoid being triggered by dew or fog.
- Movable lid and motor. The lid alternates between the two collectors. The seal during the dry period is what protects the rain sample from contamination by dust and gases.
- Wet collector. A funnel or bucket made of inert material (high-density polyethylene is the most common choice for major ions), connected to a bottle or reservoir.
- Dry collector. A shallow container, often made of borosilicate glass, which receives sedimentable material during periods without rain.
The manuals of international precipitation chemistry networks provide performance criteria. The technical manual of the Asian network EANET (Acid Deposition Monitoring Network in East Asia) requires that the lid open within one minute of the start of rain and close promptly at the end; that the sensor detect intensities of at least 0.05 mm/h and drops of 0.5 mm, regardless of wind direction; that it not be triggered by fog and dew; and that it remain at the same height as the funnel. The manual of the European program EMEP describes the heated sensor as 1 to 2 °C above air temperature and lists four sources of error for a collector with an automatic lid: aerodynamic error (the collector captures less rain than the surrounding surface), evaporation after the event, wetting (water retained on the walls), and lid failure to open during light rain.
A detail that often comes as a surprise: the sensor needs a delay to distinguish "end of rain" from "intermittent light rain." Dedicated precipitation detectors use this feature; the one in our catalog, for example, signals the end of rain with a two-minute delay. Without this, the lid would open and close dozens of times during a drizzle.
Selection criteria
Wet-only or bulk collection. If the question involves acidity, atmospheric chemistry, sources, and comparison with international networks, the standard is wet-only. The Precipitation Chemistry Programme of the World Meteorological Organization's (WMO) Global Atmosphere Watch (GAW) establishes automatic wet-only sampling as the primary objective, recognizing bulk sampling only where local conditions do not allow for another solution. If the question is the total nutrient input to a watershed, bulk sampling may suffice, as long as the method is clearly stated.
Materials in contact with the sample. For major ions (sulfate, nitrate, chloride, ammonium, sodium, potassium, calcium, magnesium), high-density polyethylene is suitable. For organic pollutants, prefer glass. For trace metals, the funnel and bottle need acid cleaning and metal-free materials. Ask the supplier what each component that touches the water is made of.
Collection area and capacity. The area defines the volume-to-depth conversion: in a collector with a 400 cm² area, 1 mm of rain corresponds to 40 mL. The reservoir capacity needs to accommodate the largest rainfall expected between collections; EANET recommends sizing it based on the maximum daily rainfall recorded at the site over the past ten years. A 10 L reservoir on a 400 cm² opening holds 250 mm, a comfortable margin for daily collection in Brazil, but one that may be insufficient for weekly collection in regions of intense rainfall.
Sensor, heating, and status output. A heated sensor, sensitive enough for light rain, is mandatory. A lid status output (dry contact) is very valuable: connected to the station's data logger, it documents when the wet collector was open and allows that time to be checked against the rain gauge records.
Power supply. In remote areas, the version with a solar panel and battery is the most practical. But keep in mind that heating the sensor and the mechanism consumes energy; check with the supplier what works in each power mode.
Where and how to install it
The site criteria are similar to those for a rain gauge, with extra precautions against contamination:
- Open, flat, grassy area. Exposed soil, dirt roads, yards, and recently tilled fields raise dust that contaminates both fractions. The EMEP manual recommends avoiding these sources within a radius of 100 m to 1 km, depending on the station's objective.
- Obstacles. No obstacle should exceed 30° relative to the rim of the collector; in other words, the horizontal distance to a tree or building should be at least twice its height.
- Local sources. EANET requires at least 100 m of distance from sources such as landfills, incinerators, parking lots, agricultural product storage, and domestic heating. For regional background stations, the distances from large sources increase to tens of kilometers.
- Height. The collector opening should be between 1.0 and 1.5 m above the ground (GAW and EANET), high enough to avoid splashes from the ground during heavy rain.
- Rain gauge alongside. GAW recommends a standard rain gauge at every collection site. EANET requires it to be in the same area, but at least 2 m from the sampler, along a line perpendicular to the prevailing wind during rainy periods, so that neither instrument "shadows" the other.
- Birds and insects. Bird droppings on the funnel or sensor invalidate samples (potassium and phosphate spike). Spike strips around the sensor and frequent inspection help prevent this.
In cities, open land is not always available, and several Brazilian studies have used building rooftops, as in the São Paulo study cited above, where the funnel was placed 1.5 m above the floor of a rooftop about 15 m high. This is an acceptable solution if it is documented and if the immediate surroundings of the rooftop are free of exhaust vents, chimneys, and construction dust.
Collection and preservation routine
Rainwater chemistry is fragile: microorganisms consume ammonium and organic acids, phosphate disappears, and pH changes within a few days if the sample is kept warm. The routine needs to be simple and repeatable:
- Frequency. GAW and EANET recommend daily collection at 9 a.m. local time. When this is not feasible, collection can be integrated over up to 7 days, provided sample integrity is maintained through refrigeration or a biocide.
- Refrigeration. This is the preferred preservation method: ideally the sample is kept refrigerated from collection to analysis (around 4 °C). Since the reservoir of a field sampler is usually not refrigerated, this reinforces the case for daily collection.
- Biocide. If refrigeration is not available, EANET recommends thymol (about 400 mg per 1,000 mL bottle, added before collection). In that case, ammonium should be determined by ion chromatography rather than by the indophenol blue method.
- Volume by weighing. Weigh the full and empty bottle; the mass gives the volume with better accuracy than a graduated cylinder, and the volume divided by the funnel area gives the collected depth.
- Cleaning and blanks. Funnels, bottles, and tubing should be washed with deionized water (and diluted acid, when the protocol requires it) until the conductivity of the final rinse falls below 1.5 µS/cm, the GAW criterion. In the Mackenzie study, the funnel was washed monthly with 1% nitric acid and rinsed at least five times, until the wash water conductivity fell below 1 µS/cm. Field blanks (deionized water that passes through the same path as the sample) reveal contamination of the system.
- Order of analyses. Measure conductivity before pH: the pH electrode releases KCl and contaminates the aliquot. Perform both measurements at a controlled temperature (25 °C is the usual reference). Rainwater has very low ionic strength, which makes the pH reading slow and prone to junction error; use an electrode suited to low-conductivity samples and check it against a dilute reference solution.
- Major ions. Routine analysis includes pH, conductivity, and the ions SO₄²⁻, NO₃⁻, Cl⁻, NH₄⁺, Na⁺, K⁺, Ca²⁺, and Mg²⁺, generally by ion chromatography. Organic acids (formate, acetate) are included when the ion balance does not close, which is common in urban areas and under the influence of burning events.
Quality control: the balances that validate the sample
Two simple checks, recommended by EANET and used in the Brazilian studies, separate a reliable dataset from a mere spreadsheet of numbers:
Ion balance (R1). By electroneutrality, the sum of cations (C, including the H⁺ from pH) should equal the sum of anions (A), in µeq/L:
R1 = (C − A) / (C + A) × 100%
EANET's criteria depend on ionic strength: ±30% when C + A is less than 50 µeq/L, ±15% between 50 and 100 µeq/L, and ±8% above 100 µeq/L.
Calculated × measured conductivity (R2). In dilute solutions, conductivity is the sum of the contributions of each ion, each with its own molar conductivity. H⁺ has the highest value (349.7 S·cm²/mol, versus 71.4 for nitrate and 73.5 for ammonium), which is why acidity carries so much weight: the H⁺ alone in a rain sample with pH 5.0 already accounts for about 3.5 µS/cm. The comparison is:
R2 = (Λcalc − Λmeas) / (Λcalc + Λmeas) × 100%
with a tolerance of ±20% for measured conductivity below 0.5 mS/m (5 µS/cm), ±13% between 0.5 and 3 mS/m, and ±9% above 3 mS/m. Samples outside these criteria are reanalyzed or flagged. A systematic imbalance often points to an unmeasured ion (bicarbonate in alkaline rain, organic acids in urban rain).
Other useful controls:
- Collection efficiency: compare the depth obtained from the sampler's volume with that of the rain gauge. Large, repeated differences point to lid failure, leakage, evaporation, or a transcription error.
- Completeness: EANET uses %TP, the percentage of total rainfall for the period that is associated with valid samples. An annual average based on samples representing only half the rainfall is not comparable to one based on 90%.
- Marine ratios: at coastal sites, sodium is used as a tracer for sea salt to separate the "non-marine" fraction of sulfate, calcium, and magnesium. When the Na⁺/Cl⁻ ratio deviates significantly from that of seawater, as occurred in downtown São Paulo, the researcher needs to justify an alternative choice of tracer.
Cross-referencing deposition with the weather station
The sampler tells you what fell; the station tells you how much, when, and from where. The GAW manual recommends that the collection site be located at a standard meteorological station or, at minimum, have measurements of temperature, wind direction, and wind speed. In practice, the combination opens up several types of analysis:
- Rainfall depth and intensity. The tipping-bucket rain gauge provides the official rainfall depth for the period, used to calculate flux and VWM, and the intensity at short intervals. Light, long rains tend to be more concentrated; the first millimeters of an event "wash" the atmosphere and carry the greatest load.
- Lid verification. By recording the lid status on the same data logger that counts the rain gauge pulses, you can see whether the collector was open throughout the entire rain event. Minutes of rain with the lid closed explain low collection efficiency.
- Wind sector during rain. Wind direction during rainy periods, measured according to the good practices described in our guide on wind measurement at the weather station, makes it possible to separate events influenced by the sea, an industrial zone, or agricultural areas. This is how potassium and ammonium are related to the sugarcane burning season, or chloride to the sea breeze.
- Air mass trajectories. Back-trajectory models use large-scale meteorological fields, but local interpretation depends on the wind, temperature, and humidity recorded at the station itself.
- Dry fraction and air conditions. The mass of sedimentable dust collected during the dry period relates to wind speed, relative humidity (dry soil raises more dust), and the number of days without rain. Researchers who already measure airborne particles can compare with the station's PM2.5 and PM10 particulate matter readings, keeping in mind that fine particles barely settle in a container.
The dry fraction: what it measures and what it does not measure
The dry container of a wet/dry sampler mainly collects large particles that settle under their own weight. The material is recovered using deionized water (or the appropriate solvent for organics), and the result can be expressed as mass per area and time, or chemically analyzed (ions, metals, minerals).
This quantity has regulatory application in Brazil. In the state of Espírito Santo, where sedimentable dust is a long-standing issue in the Greater Vitória metropolitan area, Decree No. 6076-R of 2025 set standards for sedimentable dust in g/m² per 30 days, with intermediate standards of 12, 10, 8, and 6 g/m²/30 days and a final standard of 5 g/m²/30 days, and requires, for regulatory purposes, sampling and analysis by an accredited laboratory. For research purposes, the dry container of the automatic sampler is an excellent tool for comparison and chemical characterization; if the goal is to verify legal compliance, follow the method and collector geometry required by the environmental agency.
And what the dry fraction does not measure: the dry deposition of gases and fine particles onto actual vegetation. For that, the approach is inferential, combining air concentrations with deposition velocity models that depend on wind, atmospheric stability, and the surface. One more reason to keep the weather station alongside.
Common mistakes
- Installing the sampler under trees, next to walls, or alongside dirt roads, contaminating both fractions.
- Using the depth estimated from the collector's volume as the official rainfall, instead of the value measured by the rain gauge.
- Leaving the sample for days in a warm reservoir without biocide, and then interpreting the drop in ammonium as an environmental signal.
- Measuring pH before conductivity in the same aliquot.
- Calculating arithmetic means instead of volume-weighted means.
- Not recording the lid status, leaving no way to explain low collection efficiency.
- Not running field blanks or the R1 and R2 balances.
- Comparing the "dry deposition" from the container with inferential gas fluxes as if they were the same quantity.
The sampler in our catalog
The automatic wet and dry deposition sampler (ref. Wet&Dry, manufactured by the Italian company Nesa) follows the design described in this guide: a motorized lid alternates between the wet and dry collectors according to the rain sensor. According to the manufacturer's catalog and product data sheet:
- a removable borosilicate glass dry collector, about 23 cm in diameter, 3 cm in height, and 400 cm² in area, also suitable for organic micropollutants;
- a high-density polyethylene wet collector, described by the manufacturer as compliant with WMO specifications, connected to a reservoir with a typical capacity of 10 L;
- a rain presence sensor with heating that can be switched off, and a thermostat that activates automatically below +5 °C;
- a status output for the opening of the wet and dry collectors via relay contacts, which can be connected to a data logger to document the collection periods;
- a structure of anodized aluminum and stainless steel, fully disassemblable, with a 1 m mounting pole, rated for operation from −30 to +70 °C;
- powered by the electrical grid or by a solar panel with battery, for autonomous operation.
For the researcher, the strong points are the genuine separation between fractions, the glass dry collector (which broadens the range of possible analyses), and the status output for integration with the weather station. Points to work out during quoting: the power configuration (grid or solar) and the power consumption of each version, since heating of the mechanism depends on grid power; and, since the field reservoir is not refrigerated, the daily collection routine or the use of a biocide.
Alternatives and complementary products from our catalog
- Heated capacitive precipitation detector (ref. HD2013.2, Delta OHM): indicates rain or snow (ON/OFF output), has an analog output proportional to intensity, and signals the end of rain with a two-minute delay. Useful for recording event duration at the station and for checking the lid's operation.
- Tipping-bucket rain gauge with a 400 cm² collecting surface (ref. PL400, Nesa): resolution of 0.2 mm, range of 0 to 300 mm/h, and a heated version available. Provides the official rainfall depth for calculating the deposition flux, with the same opening area as the collector.
- Two-axis ultrasonic anemometer (ref. HD52.3D, Delta OHM): wind speed and direction with no moving parts, with a built-in compass and Modbus RTU and SDI-12 outputs. The basis for analyzing wind sector during rain events.
- Dual-channel pH and conductivity benchtop meter (ref. PC1100-S, HORIBA): pH with resolution down to 0.001 and conductivity with up to four-point calibration, for the first two measurements of each sample in the laboratory. For rainwater, pair it with an electrode suited to low ionic strength.
- Conductivity benchtop meter (ref. EC1100-S, HORIBA): selectable cell constant (0.1, 1.0, or 10), useful for rainwater's low conductivity, for checking rinse water, and for the R2 balance.
- Housing with acquisition system for weather stations (ref. HD32.35, Delta OHM): a data logger in an IP66 stainless steel housing, powered by the grid or a solar panel, to bring the rain gauge, wind data, and lid status together into a single file.
For further reading, see also our guides on how to choose a data logger and on weather station telemetry.
Checklist before requesting a quote
- What is the research question: rain acidity, nitrogen input, marine influence, burning events, metals, or dust from a specific source?
- Will collection be wet-only, bulk, or separate wet and dry?
- What analyses will be performed (major ions, organics, metals), and what materials can the sample come into contact with?
- What collection frequency is feasible (daily at 9 a.m. or up to weekly), and will refrigeration or a biocide be used?
- Does the site meet the exposure criteria (30°, twice the height of obstacles, distance from sources)?
- Is there a rain gauge and anemometer at the same site, and a data logger to also record the lid status?
- What power supply is available (grid or solar)?
- How will field blanks, the R1 and R2 balances, and the completeness calculation be handled?
References
- World Meteorological Organization. Manual for the GAW Precipitation Chemistry Programme: Guidelines, Data Quality Objectives and Standard Operating Procedures. GAW Report No. 160 (WMO/TD-No. 1251). Genebra: WMO, 2004.
- Network Center for EANET. Technical Manual for Wet Deposition Monitoring in East Asia – 2010. Acid Deposition Monitoring Network in East Asia, 2010.
- EMEP/CCC. EMEP Manual for Sampling and Chemical Analysis, cap. 3: Sampling methods. Norwegian Institute for Air Research (NILU).
- Leal, T. F. M.; Fontenele, A. P. G.; Pedrotti, J. J.; Fornaro, A. Composição iônica majoritária de águas de chuva no centro da cidade de São Paulo. Química Nova, v. 27, n. 6, p. 855–861, 2004.
- Lara, L. B. L. S.; Artaxo, P.; Martinelli, L. A.; Victoria, R. L.; Camargo, P. B.; Krusche, A.; Ayers, G. P.; Ferraz, E. S. B.; Ballester, M. V. Chemical composition of rainwater and anthropogenic influences in the Piracicaba River Basin, Southeast Brazil. Atmospheric Environment, v. 35, n. 29, p. 4937–4945, 2001.
- Fornaro, A.; Gutz, I. G. R. Wet deposition and related atmospheric chemistry in the São Paulo metropolis, Brazil. Part 3: Trends in precipitation chemistry during 1983–2003. Atmospheric Environment, v. 40, n. 30, p. 5893–5901, 2006.
- Espírito Santo (Estado). Decreto nº 6076-R, de 11 de junho de 2025. Regulamenta a Política Estadual de Qualidade do Ar e estabelece padrões de qualidade do ar.
- Nesa Srl. Mod. W&D – Wet&Dry Sampler – V2 (catálogo técnico).
Instruments for this application

DeltaOhm - GHM GroupHD2013.2
HD2013.2 – Rain Detector

NESASRLPL400
PL400 Rain Gauge with 400cm2 Collecting Surface

DeltaOhm - GHM GroupHD52.3D
HD52.3D Series – 2-Axis Ultrasonic Anemometer

HORIBA ScientificPC1100-S
LAQUA PC1100 Dual Channel pH/ORP/Cond/TDS/Res/Sal/Temp Benchtop Meter

HORIBA ScientificEC1100-S
LAQUA EC1100 Conductivity/TDS/Res/Sal/Temp Benchtop Meter

DeltaOhm - GHM GroupHD32.35