Particulate matter (PM2.5 and PM10) at the weather station: how to measure with a laser optical sensor, correct for humidity and use the data in research

A teaching guide to PM1, PM2.5 and PM10: definitions and Brazil's CONAMA 506/2024 standards, the light-scattering principle, humidity correction (κ-Köhler), collocation and US EPA targets, installation, data use and quality control.

By HenSistemas · · 15 min read

Particulate matter transmitter in a white enclosure with hooded air inlets and an M12 connector at the base

Of all the variables a weather station can log, particulate matter is the one that brings meteorology closest to public health. The concentration of fine particles in the air of a city, a campus or a farming area depends directly on wind, mixing-layer height, temperature inversions and rain. That is why a PM2.5 series logged next to the anemometer, the thermo-hygrometer and the rain gauge is worth far more than the same series on its own. But the laser optical sensor that made this measurement affordable has physical limitations that must be understood before drawing any conclusion: it does not weigh particles, it "sees" them, and air humidity changes what it sees.

This guide is for lecturers, graduate students and researchers at public universities and institutes who want to add PM1, PM2.5 and PM10 to their experimental site. We explain the legal definitions, the light-scattering principle, why relative humidity is the main interferent, how to choose, install and collocate the sensor, how to run quality control and how to use the data in research. The link to the weather station is direct: see also our articles on air temperature and humidity at the weather station and on atmospheric visibility, which also depends on aerosols.

What PM2.5 and PM10 are

Particulate matter is the set of solid and liquid particles suspended in air: dust, soot, smoke, secondary aerosols (sulphate, nitrate, organics), sea salt, pollen. It is classified by equivalent aerodynamic diameter, i.e. the diameter of a sphere of density 1 g/cm³ with the same settling velocity as the real particle. Brazil's CONAMA Resolution 506/2024 defines:

  • PM10: particles with an equivalent aerodynamic cut-off diameter of 10 µm;
  • PM2.5: particles with a cut-off diameter of 2.5 µm;
  • TSP: total suspended particles, with a 50 µm cut-off.

The fine fraction (PM2.5) reaches the alveoli and comes mainly from combustion (vehicles, industry, biomass burning) and from chemical reactions in the atmosphere. The coarse fraction (between 2.5 and 10 µm) comes mainly from resuspended dust, bare soil, construction and sea salt. The PM2.5/PM10 ratio is therefore a first indicator of origin: high values suggest combustion or secondary aerosol; low values suggest mineral dust.

Brazilian standards and the WHO reference

CONAMA Resolution 506 of 5 July 2024 sets standards in stages. Values in µg/m³ (PI = interim standard; PF = final standard) are:

Pollutant and period PI-1 PI-2 PI-3 PI-4 PF (WHO 2021)
PM2.5, 24 hours 60 50 37 25 15
PM2.5, annual mean 20 17 15 10 5
PM10, 24 hours 120 100 75 50 45
PM10, annual mean 40 35 30 20 15

The second interim standard (PI-2) has applied since 1 January 2025; PI-3 comes into force on 1 January 2033 and PI-4 on 1 January 2044. The final standard (PF) matches the World Health Organization guideline values published in 2021. The resolution sets 25 °C and 1013.2 hPa as reference conditions.

A key teaching point: article 5 of the same resolution requires compliance to be verified with reference or equivalent methods. A low-cost optical sensor, however good, produces indicative data, useful for research, dense networks and alerts, but it does not replace the official station for legal purposes. Making this clear in the methodology avoids trouble with examiners and peer reviewers.

How the optical sensor measures

Reference methods for PM are gravimetric: a sampler with a size-selective inlet collects particles on a filter for 24 h, and the filter is weighed on a microgram balance in a temperature- and humidity-controlled room. Equivalent automatic monitors use beta attenuation or an oscillating microbalance. They are expensive instruments that require a shelter, mains power and an operator.

The optical sensor takes another route. A small fan draws air into a dark chamber where a laser beam crosses the flow. Each particle that crosses the beam scatters light, and a photodetector records pulses whose intensity grows with particle size (Mie scattering regime, for particles of a size comparable to the wavelength). The sensor counts particles by size class and converts counts into mass assuming spherical particles of fixed density:

M = Σ Nᵢ · (π/6) · Dᵢ³ · ρₚ

where Nᵢ is the number of particles per volume in class i, Dᵢ the optical diameter of the class and ρₚ the assumed density. Three physical limitations follow:

  1. Optical diameter is not aerodynamic diameter. The size "seen" depends on the particle's refractive index and shape. Soot, mineral dust and sulphate droplets of the same size scatter light differently.
  2. Density is assumed. If the local aerosol is denser or lighter than the factory value, mass is biased in the same proportion.
  3. There is a lower size detection limit. Particles smaller than a few hundred nanometres scatter almost no light and are not counted. At the other end, coarse particles are few, and the efficiency with which they enter the chamber drops with size; that is why many low-cost optical sensors represent PM2.5 well and underestimate the coarse fraction.

Relative humidity: interferent number one

Many atmospheric aerosol particles are hygroscopic: they take up water and grow as relative humidity rises. The optical sensor measures the wet particle and therefore overestimates the dry mass, which is what the gravimetric method (with a conditioned filter) reports. κ-Köhler theory describes this growth with a single hygroscopicity parameter, κ. Crilley et al. (2018) used it to derive a mass correction factor:

C = 1 + (κ / ρₚ) · aw / (1 − aw), with aw = RH/100 and ρₚ in g/cm³

PM_dry = PM_sensor / C

With ρₚ = 1.65 g/cm³ (the density used in that work) and an illustrative κ of 0.3, the factor is about 1.18 at 50% RH, 1.42 at 70%, 1.73 at 80% and 2.64 at 90%. The lesson is clear: below 60–70% RH the effect is moderate; above 80–85% it explodes. In Brazil, where humid nights and early mornings are the rule across much of the country, ignoring humidity can artificially inflate daily means precisely in the hours when the mixing layer is lowest.

Three practical consequences:

  • Log RH alongside PM, preferably with a good humidity sensor inside a radiation shield on the same mast. The PM sensor's internal RH, where available, may be affected by the instrument's own heating.
  • Flag data with RH above 85% and treat fog and rain separately. In fog, water droplets are counted as "particles".
  • Do not apply a generic κ without validation. κ depends on the local aerosol composition. The safest route is collocation with a reference instrument, which lets you fit your own correction model.

Collocation and calibration

Optical sensors leave the factory calibrated with a test aerosol that rarely resembles the aerosol at your site. Good practice is to install the sensor next to a reference station (in Brazil, the state monitoring networks) for a few weeks, preferably covering dry and humid periods, and fit a regression such as:

PM_ref = a · PM_sensor + b · RH + c

This is the form used, for example, by Barkjohn et al. (2021) to correct a large network of optical sensors in the United States. To evaluate the sensor, the US Environmental Protection Agency (US EPA, 2021) proposes the following targets for PM2.5 as 24-hour averages in a field test protocol:

  • precision among identical units: standard deviation ≤ 5 µg/m³ or coefficient of variation ≤ 30%;
  • bias: slope of 1.0 ± 0.35 and intercept between −5 and +5 µg/m³;
  • linearity: R² ≥ 0.70;
  • error: RMSE ≤ 7 µg/m³ or NRMSE ≤ 30%.

The same document recommends that at least one test day have a 24-h PM2.5 mean of 25 µg/m³ or more, and that meteorological effects on performance be examined graphically. These are excellent targets for a methodology chapter. The World Meteorological Organization has also published a review on using these sensors to measure atmospheric composition (WMO-No. 1215), with guidance on calibration and validation.

In Brazil, WRI Brasil (2021) used low-cost PM2.5 sensors in fixed and mobile mode in the city of São Paulo to assess the "Car-Free Friday" initiative, and the resulting practice note is a good read on methodology, lessons learned and deployment best practice.

How to choose the sensor

  • Fractions measured. PM2.5 is the minimum; PM1 helps isolate combustion; PM10 is useful, but remember the coarse-fraction limitation.
  • Range. During biomass-burning episodes, hourly means of hundreds of µg/m³ occur in several regions of Brazil. Choose a range that does not saturate in these events.
  • Integration with the station. An RS485 Modbus-RTU or analog output lets you connect the sensor to the station logger and have PM on the same clock as the meteorological variables.
  • Laser and fan lifetime. Cyclic measurement modes greatly extend service life.
  • Power consumption. At solar-powered sites, average consumption sizes the battery.
  • Inlet protection against rain, insects and UV radiation.

The reference sensor in this guide

The main instrument in this guide is the Delta OHM particulate matter transmitter (PMsense). According to the product sheet and the manufacturer's documentation:

  • laser light scattering principle, measuring PM1.0, PM2.5 and PM10;
  • range of 0 to 1000 µg/m³ for each fraction and particle detection from 0.3 to 10 µm;
  • linearity error < 5%, repeatability < 3% and temperature drift < 0.01 µg/m³/°C;
  • RS485 output with Modbus-RTU (or a proprietary ASCII protocol) and, as an option, two configurable analog outputs of 4–20 mA, 0–5 V or 0–10 V;
  • 7 to 30 VDC supply; 25 mA at 24 VDC while measuring and 4 mA on standby in cyclic mode;
  • cyclic mode (default, 5-minute interval and a sensor life of about 5 years) or continuous mode (1 measurement per second and more than 10,000 h of life). According to the manual, in cyclic mode the measurement becomes available 70 s after the sensor is switched on;
  • selectable averages of 10 s, 60 s or 15 min;
  • operation from −20 to +70 °C and 500 to 1500 hPa; IP53 polycarbonate housing with a rain-proof, UV-resistant air inlet filter; 120 × 94 × 71 mm, 330 g; bracket with U-bolt for 35 to 44 mm masts;
  • option with an integrated NDIR CO₂ sensor (0 to 5000 ppm).

Its strength for research is integration: via Modbus-RTU it plugs into the weather station logger and records PM alongside wind, temperature, humidity, pressure and rain, on the same clock and telemetry system.

Alternatives and complements from the catalogue

  • Weather station data logger with 4G (HD33MT.4): RS485 Modbus-RTU port configurable as master, SDI-12 master port, analog and counter inputs, and data upload over mobile networks. It is the link that merges the PM sensor and the meteorological sensors into a single series.
  • Relative humidity and temperature sensor with SDI-12 output (SRH1A): capacitive sensor for installation in a radiation shield, compatible with SDI-12 loggers. It provides the RH needed to flag and correct hygroscopic growth.
  • Air quality station with laser particle sensor and Wi-Fi (TSI BlueSky 8145): measures PM1.0, PM2.5, PM4.0 and PM10 simultaneously, with a PM sensor factory-calibrated by TSI and the option of a custom calibration factor, plus temperature, humidity and pressure; sends data to the cloud over Wi-Fi and keeps a copy on an SD card. Practical for dense campus networks.
  • Solar-powered stand-alone air quality station (Ostara): PM1, PM2.5 and PM10 from 0 to 1000 µg/m³, integrated temperature, humidity and pressure, interface for an ultrasonic anemometer, optional gas sensors, 2G/3G/4G transmission, IP66 enclosure and photovoltaic power. Suited to remote sites.
  • Air quality monitoring station with GPRS (HS-EMQA): PM2.5 (0–1000 µg/m³) and PM10 (0–2000 µg/m³), plus CO, SO₂, NO₂, O₃, temperature and humidity, with GPRS transmission and GPS. Useful when the question also involves gases.
  • Particulate matter and noise sensor (EIC-28PM2523): PM2.5 and PM10 from 0 to 500 µg/m³ and sound level from 30 to 130 dB(A), with RS485, 4–20 mA or 0–5 V output and IP54 protection. A simple option for urban studies combining pollution and noise.

Installation

  1. Define the spatial scale of the question. Pedestrian exposure, urban background, roadside or farmland call for different locations. Record this choice in the methodology.
  2. Inlet height. As a reference, US EPA criteria (40 CFR Part 58, Appendix E) require an inlet between 2 and 15 m above ground for neighbourhood or larger scales, and between 2 and 7 m for microscale.
  3. Clearance from obstacles. Under the same criteria, the inlet must be at least 1 m from walls and supporting structures, and at least 10 m (preferably 20 m) from the drip line of trees.
  4. Local sources. Keep the sensor away from exhaust vents, barbecues, generators, dirt roads and material stockpiles, unless the source is the object of study.
  5. Same mast, same clock. Install the thermo-hygrometer and anemometer nearby and use the same logger, in UTC.
  6. Free and protected inlet. Follow the manufacturer's mounting orientation; the inlet must not be obstructed by tape, cables or insect nests.

Using the data in research

  • Dispersion meteorology. The ventilation coefficient, the product of wind speed and mixing-layer height, explains much of the daily variation. Calm early mornings with temperature inversions concentrate pollutants; cold fronts and rain clean the atmosphere (wet scavenging). Having PM and meteorology on the same logger makes these analyses straightforward.
  • Biomass burning. In the dry season, smoke can raise PM2.5 over large areas of Brazil's Centre-West and North and, through transport, in other regions. A high PM2.5/PM10 ratio and reduced visibility help identify these episodes.
  • Agriculture. Dust from tillage, harvesting and rural roads affects workers, settles on leaves and reduces the radiation reaching the canopy.
  • Urban climate and health. PM2.5 series feed epidemiological and heat-island studies. Andrade et al. (2017) reviewed thirty years of air quality in the São Paulo Metropolitan Area and are a good starting point.
  • Dense networks. Several sensors collocated and then spread across a campus allow the study of spatial gradients that a single official station cannot resolve.

Quality control

  • Collocation before and after. Compare with a reference station at the start and end of the campaign; the difference between the two regressions estimates drift.
  • Inter-unit comparison. Before spreading the sensors, run them side by side for a few days and compute the coefficient of variation among them.
  • Humidity filter. Flag data with RH ≥ 85%, fog and rain.
  • Physical consistency. PM1 ≤ PM2.5 ≤ PM10, always. Inversions point to a communication or configuration error.
  • Constant values, zeros and saturation. Identical readings for hours, persistent zeros or full-scale values indicate a stopped fan, a blocked inlet or a communication failure.
  • Completeness. Compute 24-hour means only with a minimum fraction of valid data (75% of hours is a common criterion) and report the criterion used.
  • Maintenance. The reference sensor's manual asks for periodic checks of the cleanliness of the air inlet and outlet grids, using a soft dry cloth or one slightly dampened with clean water, without aggressive cleaning agents.
  • Metadata. Serial number, firmware, measurement mode, averaging interval, inlet height, distances to obstacles and sources, correction equation and collocation period.

Checklist before requesting a quote

  1. What is the research question and the spatial scale (exposure, urban background, specific source, rural area)?
  2. Are PM2.5 and PM10 enough, or are gases also needed?
  3. Will the sensor connect to the station logger (Modbus-RTU, analog) or work as a stand-alone station with cloud access?
  4. Is there mains power, or will solar panel and battery be needed?
  5. Is there a good relative humidity sensor on the same mast?
  6. Is there a reference station nearby for collocation?
  7. What averaging interval will be used, and how will 24-h and annual means be computed?

References

  • Andrade, M. F. et al. Air quality in the megacity of São Paulo: evolution over the last 30 years and future perspectives. Atmospheric Environment, v. 159, p. 66–82, 2017.
  • Barkjohn, K. K.; Gantt, B.; Clements, A. L. Development and application of a United States-wide correction for PM2.5 data collected with the PurpleAir sensor. Atmospheric Measurement Techniques, v. 14, p. 4617–4637, 2021.
  • Brazil. National Environment Council. CONAMA Resolution No. 506 of 5 July 2024. Sets national air quality standards. Diário Oficial da União, 9 July 2024.
  • Crilley, L. R. et al. Evaluation of a low-cost optical particle counter (Alphasense OPC-N2) for ambient air monitoring. Atmospheric Measurement Techniques, v. 11, p. 709–720, 2018.
  • Petters, M. D.; Kreidenweis, S. M. A single parameter representation of hygroscopic growth and cloud condensation nucleus activity. Atmospheric Chemistry and Physics, v. 7, p. 1961–1971, 2007.
  • U.S. EPA. Performance testing protocols, metrics, and target values for fine particulate matter air sensors. EPA/600/R-20/280, 2021.
  • U.S. EPA. 40 CFR Part 58, Appendix E: Probe and monitoring path siting criteria for ambient air quality monitoring.
  • WMO. An update on low-cost sensors for the measurement of atmospheric composition. WMO-No. 1215. Geneva, 2021.
  • WRI Brasil. Measuring PM2.5 with low-cost sensors. Practice note, 2021.
  • World Health Organization. WHO global air quality guidelines: particulate matter (PM2.5 and PM10), ozone, nitrogen dioxide, sulfur dioxide and carbon monoxide. Geneva, 2021.
  • Delta OHM / Senseca. PMsense: datasheet and operating manual.