Agriculture · Technical guides

Portable chlorophyll meters in research: how to measure, calibrate and interpret leaf readings

A teaching guide to the portable chlorophyll meter: what the SPAD index measures (650 and 940 nm), why to calibrate against acetone or DMSO extraction, the link to leaf nitrogen, sampling protocols used in Brazil for maize and coffee, the method's limits, data logging and a researcher checklist.

By HenSistemas · · 15 min read

Portable chlorophyll meter being used on a maize leaf in the field

Few field measurements are as fast as a portable chlorophyll meter reading: clamp the leaf, close the measuring head and, in about two seconds, a number appears on the display. That speed explains why the instrument is so common in crop science, plant physiology, soil science, ecology and forestry research in Brazil. It also explains a frequent problem: the number is easy to get, but turning it into scientific data takes method.

This guide explains, in a teaching format, what the SPAD index actually measures, how it relates to laboratory-extracted chlorophyll and to leaf nitrogen, which sampling protocols Brazilian studies use, where the method's limits lie, and what to check before buying the instrument.

What the meter measures: transmittance in two light bands

The meter does not "see" chlorophyll directly. It measures how much light passes through the leaf at two wavelengths. According to the manufacturer's manual, a red LED (peak at about 650 nm) and an infrared LED (peak at about 940 nm) fire in sequence, and the light transmitted by the leaf reaches a silicon photodiode.

The choice of bands follows a simple logic. Chlorophyll absorbs strongly in the blue and red and hardly at all in the near infrared. The red band therefore carries the pigment information, while the infrared band acts as an internal reference that discounts effects unrelated to chlorophyll, such as leaf thickness and structure. Before each session, the meter is calibrated by closing the head with no sample, and the displayed value comes from comparing the leaf's transmittance with that reference reading.

The result is the SPAD value: a dimensionless index defined by the manufacturer and correlated with the amount of chlorophyll per unit leaf area. It is not a concentration in µg/cm² or mg/g. In Brazilian literature it also appears as the relative chlorophyll index (índice relativo de clorofila, IRC).

SPAD is not a concentration: why calibrate against extraction

The relationship between the index and extracted chlorophyll is not linear. Working with soybean and maize, Markwell, Osterman and Mitchell (1995) described it with an exponential equation (r² = 0.94) and warned that the exact curve varies from one meter to another. Uddling et al. (2007) confirmed the non-linearity in birch, wheat and potato: the relationship was strong in birch and wheat (r² close to 0.9) but weak in potato (r² close to 0.5). The authors attribute the curvature mainly to the uneven distribution of chlorophyll within the leaf.

For forest species, Coste et al. (2010) calibrated the index for 13 tree species of the tropical rainforest in French Guiana. A general model explained 89% of the variation; species-specific models reached 96%. The lesson for ecology and forestry is clear: a generic model may serve broad comparisons, but species-specific calibration improves the estimate.

In practice, when the goal is to talk about chlorophyll (rather than just "greenness"), the calibration protocol is:

  1. Read the index exactly where the leaf will be sampled, ideally marking the spot.
  2. Punch leaf discs of known area from that same spot.
  3. Extract the pigments in 80% acetone (the classic Arnon method, 1949) or in dimethyl sulfoxide (DMSO), which avoids grinding (Hiscox and Israelstam, 1979).
  4. Read on a spectrophotometer and use up-to-date equations. Porra et al. (1989) showed that Arnon's original coefficients overestimate concentrations, especially of chlorophyll b; the equations of Porra et al. (1989) or Lichtenthaler (1987) are preferable.
  5. Express chlorophyll per unit area (µg/cm² or µmol/m²), consistent with what the meter "sees", and fit a non-linear model. Report the equation, r², number of leaves and the range of values covered.

One technical detail deserves attention: the accuracy stated by the manufacturer (±1.0 SPAD unit) applies to readings between 0 and 50. Above 50, the display shows an asterisk and the manual warns that accuracy is not guaranteed. Many crops sit above that value, such as maize at silking and coffee for much of its cycle. This does not invalidate the readings, but it reinforces the need for your own calibration and for comparing treatments on the same instrument.

Chlorophyll and nitrogen: a useful but indirect link

A large share of leaf nitrogen sits in chloroplast proteins, which is why the index tracks the plant's nitrogen status. The manufacturer's catalogue shows a linear relationship between the index and leaf-blade N concentration (r = 0.908; n = 68). Brazilian reviews (Argenta, Silva and Bortolini, 2001; Rambo et al., 2004) highlight one advantage and one limitation. The advantage is that the index is little affected by luxury nitrogen uptake as nitrate, which raises leaf N without raising yield. The limitation is that readings also respond to hybrid or cultivar, growth stage, leaf thickness, water stress, other nutrients, pests and light.

That is why there are two ways to use it:

  • Critical values. In maize, Argenta et al. (2001), in the Revista Brasileira de Fisiologia Vegetal, found a positive correlation between the index and extracted chlorophyll, but the relationship with leaf N only became consistent from the 10–11 leaf stage. In a review, Argenta, Silva and Bortolini (2001) compiled critical levels of 40–42 for rice (panicle differentiation), 41–42 for wheat (flag leaf) and 58–62 for maize (silking). These numbers hold for the conditions in which they were obtained.
  • Sufficiency index (reference strip). Divide the reading from the area being assessed by the reading from a strip of the same cultivar with non-limiting nitrogen. Varvel, Schepers and Francis (1997) used a 95% threshold to trigger side-dressing in maize and found that, below 90% at V8, late application no longer restored maximum yield.

The sufficiency index is not infallible. In a high-yield maize field in the Brazilian Cerrado, Hurtado et al. (2011), with researchers from Embrapa Maize and Sorghum, found that a fixed 0.95 threshold throughout the cycle called for up to 180 kg/ha of side-dressed N, above the doses for maximum technical (144 kg/ha) and economic (105 kg/ha) efficiency. The meter separated contrasting nitrogen conditions well but lacked the sensitivity to decide on small extra doses. The authors' conclusion is the best rule of thumb: calibrate locally.

What Brazilian research has shown

  • Maize. The most studied crop in the country, with critical values by growth stage, reading protocols and sufficiency-index trials in southern Brazil and the Cerrado.
  • Coffee. Godoy et al. (2008), at UNESP Botucatu, monitored fertigated Catuaí Vermelho coffee. The index rose linearly with N rates and correlated with yield, but not with leaf N. Values in the most productive plants fell over the cycle, from 81.5–83.2 at flowering to 61.7–62.7 at maturation. In other words, there is no single value for the whole year. Reis et al. (2006) also reported a correlation between the index, leaf N and coffee yield, and Torres Netto et al. (2005) between the index, photosynthetic pigments and nitrogen.
  • Sugarcane. Amaral and Molin (2011), at ESALQ/USP, found a low correlation between the index and leaf N. For variable-rate nitrogen recommendations, an active canopy optical sensor was more consistent with yield. It is a good example of a crop where the leaf meter should be used with caution.
  • Native and forest species. The work of Coste et al. (2010) shows that the method applies to tropical trees, as long as it is calibrated.

Sampling protocol: where data quality is won or lost

The manufacturer's manual and published protocols agree on the main points:

  • Calibrate at power-on. The no-sample calibration is required every time the meter is switched on.
  • Standardize the leaf. In maize, Hurtado et al. (2011), following Argenta et al. (2001), read the last fully expanded leaf until flowering and the leaf below and opposite the ear afterwards. Readings were taken in the middle third, on both sides of the blade, avoiding the midrib and margins. In coffee, Godoy et al. (2008) used the third or fourth leaf pair, two points per leaf (one on each side of the midrib) and 16 readings per plant.
  • Avoid veins. The leaf must fully cover the receiving window (2 × 3 mm measuring area). The manual says not to measure thick parts such as veins; on leaves with many fine veins, take several readings and use the average.
  • Repeat and average. The meter itself averages the values in memory. Set the number of readings per leaf, leaves per plant and plants per plot in the project, and keep it the same across all assessments.
  • Fix the position. The depth stop, adjustable from 0 to 6 mm from the centre line, keeps the same distance from the leaf edge on every leaf.
  • Control time of day and light. Martínez and Guiamet (2004) found that, in well-watered field maize, the index varied by up to 4 units over the day. A drop in wheat leaf relative water content (from 94% to 87.5%) raised readings by 2 to 3 units. Hoel and Solhaug (1998) recorded, in wheat, the lowest readings at midday and the highest at dawn and dusk. Choose a fixed time window (Godoy et al. used 9 to 10 a.m.) and, in direct sun, shade the meter with your body, as the manual recommends.
  • Watch out for moisture. Water droplets or dirt on the emitting and receiving windows spoil the measurement. Avoid measuring leaves wet with dew or rain, and wipe the windows with a soft, dry cloth.
  • Use reference plots. For nitrogen diagnosis, include a non-limiting N strip in the design so you can calculate the sufficiency index.

Data recording: internal memory or built-in logger

The standard model stores 30 readings, shows a trend graph on the display and calculates the average. There is an important operational detail: according to the manual, memory is erased when the meter is switched off. With the non-logging version, data must therefore go into a field book, tablet or spreadsheet during collection.

The version with a built-in data logger, sold by Spectrum Technologies, stores 4,096 readings (or 1,488 when paired with GPS coordinates). It has an RS-232 serial port for connection to a GPS receiver with serial output, and data are downloaded with the FieldScout software, which the distributor provides free of charge. For field mapping or large trials, this version saves time and reduces transcription errors.

Care and maintenance

  • Reading checker. The meter ships with a reading checker bearing the same serial number. Use it periodically, in check mode, indoors and out of direct sunlight. The average should fall within the range printed on the checker; if it does not, clean the windows and repeat the test. If the deviation persists, contact technical service.
  • Several meters in one project. According to the manual, different units give slightly different results on the same leaf. The manufacturer provides a compensation factor (from −9.9 to +9.9) to match meters to a master unit. This factor must not be used to correct deviations on the reading checker.
  • Water and cleaning. The meter is splash-resistant (IPX4) and can be used in the rain, but must not be immersed or washed. Clean it with a soft, dry cloth; the manufacturer advises against alcohol or chemicals on the housing.
  • Storage. Avoid heat, sun and humidity (for example, inside a closed car) and store with a desiccant. For periods longer than two weeks, remove the batteries.
  • Batteries. Two AA alkaline batteries last more than 20,000 measurements, according to the manufacturer.

An example from the catalogue

In the HenSistemas catalogue, the Konica Minolta portable chlorophyll meter (model SPAD-502Plus, distributed by Spectrum Technologies) brings together the features described in this guide. According to the manufacturer's data sheet:

  • Measurement by optical density difference at two wavelengths (about 650 and 940 nm), with a 2 × 3 mm measuring area.
  • Leaves up to 1.2 mm thick and insertion depth up to 12 mm.
  • Display range of −9.9 to 199.9 SPAD units.
  • Accuracy of ±1.0 unit (between 0 and 50), repeatability of ±0.3, reproducibility of ±0.5 and temperature drift of ±0.04 unit/°C.
  • Operation from 0 to 50 °C, at up to 85% relative humidity (at 35 °C), non-condensing.
  • Minimum interval of about 2 s between measurements, 78 × 164 × 49 mm and 200 g without batteries.
  • Supplied with depth stop, strap, case and reading checker.

The line remains in Konica Minolta's current portfolio, which lists it as the successor to the previous model, discontinued in 2010. When quoting, the HenSistemas team confirms the version (with or without data logger) and the accessories included.

Related products

Checklist before requesting a quote

  1. Is the goal to diagnose nitrogen, compare treatments or estimate chlorophyll in absolute units? For the last one, plan a calibration with laboratory extraction.
  2. Do the leaves of the species studied cover the 2 × 3 mm window and measure no more than 1.2 mm thick? Very narrow, needle-like or thick leaves call for a preliminary test.
  3. How many readings per campaign? If it is hundreds a day, or if you will map with GPS, the built-in logger version usually pays off.
  4. How many meters will the group use? With more than one, plan inter-unit standardization.
  5. Do you have a spectrophotometer, solvent (80% acetone or DMSO) and a leaf punch for calibration?
  6. Does the experimental design include reference strips with non-limiting nitrogen?
  7. Who will run the periodic reading-checker test, and where will the meter be stored?
  8. Will the purchase go through a research grant or a public procurement process? If so, ask for the technical specification for the terms of reference.

A well-defined protocol before the first reading avoids the most common problem with this kind of instrument: series of numbers that cannot be compared across dates, teams or seasons. If you would like help matching the instrument to your research protocol, talk to the HenSistemas team.

References cited

  • AMARAL, L. R.; MOLIN, J. P. Pesquisa Agropecuária Brasileira, v. 46, n. 12, p. 1633-1642, 2011. DOI: 10.1590/S0100-204X2011001200008.
  • ARGENTA, G. et al. Revista Brasileira de Fisiologia Vegetal, v. 13, n. 2, p. 158-167, 2001. DOI: 10.1590/S0103-31312001000200005.
  • ARGENTA, G.; SILVA, P. R. F.; BORTOLINI, C. G. Clorofila na folha como indicador do nível de nitrogênio em cereais. Ciência Rural, v. 31, n. 4, p. 715-722, 2001.
  • ARNON, D. I. Plant Physiology, v. 24, n. 1, p. 1-15, 1949.
  • COSTE, S. et al. Annals of Forest Science, v. 67, 2010. DOI: 10.1051/forest/2010020.
  • GODOY, L. J. G. et al. Índice relativo de clorofila e o estado nutricional em nitrogênio durante o ciclo do cafeeiro fertirrigado. Revista Brasileira de Ciência do Solo, v. 32, n. 1, p. 217-226, 2008.
  • HISCOX, J. D.; ISRAELSTAM, G. F. Canadian Journal of Botany, v. 57, p. 1332-1334, 1979. DOI: 10.1139/b79-163.
  • HOEL, B. O.; SOLHAUG, K. A. Annals of Botany, v. 82, p. 389-392, 1998.
  • HURTADO, S. M. C. et al. Clorofilômetro no ajuste da adubação nitrogenada em cobertura para o milho de alta produtividade. Ciência Rural, v. 41, n. 6, p. 1011-1017, 2011.
  • LICHTENTHALER, H. K. Methods in Enzymology, v. 148, p. 350-382, 1987.
  • MARKWELL, J.; OSTERMAN, J. C.; MITCHELL, J. L. Photosynthesis Research, v. 46, p. 467-472, 1995. DOI: 10.1007/BF00032301.
  • MARTÍNEZ, D. E.; GUIAMET, J. J. Agronomie, v. 24, p. 41-46, 2004. DOI: 10.1051/agro:2003060.
  • PORRA, R. J.; THOMPSON, W. A.; KRIEDEMANN, P. E. Biochimica et Biophysica Acta, v. 975, p. 384-394, 1989. DOI: 10.1016/S0005-2728(89)80347-0.
  • RAMBO, L. et al. Parâmetros de planta para aprimorar o manejo da adubação nitrogenada de cobertura em milho. Ciência Rural, v. 34, n. 5, 2004.
  • REIS, A. R. et al. Diagnóstico da exigência do cafeeiro em nitrogênio pela utilização do medidor portátil de clorofila. Bragantia, v. 65, n. 1, p. 163-171, 2006.
  • TORRES NETTO, A. et al. Scientia Horticulturae, v. 104, p. 199-209, 2005. DOI: 10.1016/j.scienta.2004.08.013.
  • UDDLING, J. et al. Photosynthesis Research, v. 91, p. 37-46, 2007. DOI: 10.1007/s11120-006-9077-5.
  • VARVEL, G. E.; SCHEPERS, J. S.; FRANCIS, D. D. Soil Science Society of America Journal, v. 61, p. 1233-1239, 1997. DOI: 10.2136/sssaj1997.03615995006100040032x.