The HS148000 is a gas analyzer that doubles as a data collector, quickly recording and analyzing gas concentrations in scientific research, university settings and complex working environments.
It detects up to 18 gases at the same time, chosen from more than 45 gas types.
Price on request
No sign-up. You review the list and send it when you are ready.
Torn between two models?
Ask on WhatsAppSaves this product and the next ones you view in this browser, up to eight, for 30 days. You can disable it under “History” in the footer.
History is enabled in this browser. History
Each entry lists the standard range, the other ranges available and the resolution.
For other gases or measuring ranges, please contact us.
| Property | Value |
|---|---|
| Sampling | Industrial pump gas sampling with ten selectable pumping flow rate levels |
| Sensors | Sensors of different principles can be combined in any number, up to 18 sensors. Combustible: catalytic combustion or infrared; oxygen: electrochemical or zirconia; toxic: electrochemical; CO2: infrared; VOC: PID (photoionization); CH4: catalytic combustion or infrared; He: thermal conductivity; temperature and humidity sensors and pressure sensors (optional); other sensor principles can be customized |
| Measuring range | 0-1, 10, 100, 1,000, 5,000, 50,000, 100,000 PPM; 0-100mg/L, 200mg/L; 0-100%LEL; 0-10%, 20%, 50%, 99.999%, 100%Vol (range is optional and can be customized) |
| Resolution | 0.001 PPM (0-1 PPM); 0.01 PPM or 0.001 PPM (0-10 PPM); 0.1 PPM or 0.01 PPM (0-100 PPM); 1 PPM or 0.1 PPM (0-1000 PPM); 1 PPM (ranges above 1000 PPM); 0.01mg/L (0-200%mg/L); 0.1%LEL (0-100%LEL); 0.01% (0-100%VOL) |
| Precision | ≤±2%F.S (higher precision available on request) |
| Temperature and humidity | Temperature: -20-50℃, resolution: 0.1℃; humidity: 0-99%RH, resolution: 1%RH |
| Pressure | 0.5 – 2.0 bar (optional) |
| Screen | 3.5-inch HD LCD screen, 320* 480 resolution |
| Display | Formula, concentration data, unit, curve, battery level, time, temperature, humidity, pressure (optional), pump status, wireless status, printer status and screenshot status |
| Alarm | Buzzer, red LED and alarm status on screen; fault alarm, abnormal pump alarm and low battery alarm |
| Data storage | Independent SD memory card, Max.32G; automatic recording of measurement data with a configurable logging interval |
| Signal output / data download | RS232, RS485; data downloadable to a computer as TXT files over USB; can be supplied with PC software to download, store, analyze and print the data |
| Printing (optional) | Internal portable printer or external Bluetooth printer |
| Print content (optional) | Text, graphics, one-dimensional codes, QR codes, curves, characters, etc. |
| Charger | Standard Mini-USB charging port; 13V/2A dedicated charger (internal printer) or 5V/2A standard charger (external printer) |
| Battery | High-capacity rechargeable lithium polymer battery: 5,400mAh/11.1V (internal printer) or 10,000mAh/3.7V (external Bluetooth printer); 32h continuous with a single gas and 24h continuous with 6 gases |
| Language | English |
| Humidity | 0-95%RH (non-condensing); optional filter device for high humidity |
| Temperature | -20℃~+50℃; optional gas sampling gun that measures gas concentration at gas temperatures up to 1200 degrees |
| Explosion-proof | Ex ic IIC T3 Gc |
| Degree of protection | IP65 |
| Weight | 3700g |
| Warranty | 12 months |